S15 · Problem set 9
Problem set 9: ultrasound echoes, attenuation and resolution, MRI frequencies, contrast and encoding, tracer decay, stimulation thresholds, pacing energy and defibrillation, dialysis and ventilation, shock and leakage currents, predictive values, and power budgets and reliability
These problems cover lessons 40, 41, 43 and 45 to 48: reflections at tissue boundaries, echo depths, attenuation and frame rates, matching layers, Larmor frequencies, polarization, spin-echo contrast, gradients and slices, tracer decay, strength–duration thresholds, pacemaker and defibrillator energy, dialysis clearance and Kt/V, body and leakage currents, sensitivity, specificity and predictive values, ground-potential differences, duty-cycled power, radio energy, SAR and reliability.
Check
Ultrasound crosses from 1.38 MRayl into 1.75 MRayl; another echo returns after 150 µs.
(a)
Reflected intensity fraction?
(b)
Transmitted fraction?
(c)
Depth of the second reflector?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
Check
Tissue at 0.5 dB/(cm·MHz), a 3 MHz 3-cycle pulse, 70 dB of dynamic range.
(a)
Deepest imageable depth?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
(b)
Axial resolution?
(c)
Frame rate for 128 lines to that depth?
Check
A 1.5 T scanner at 300 K with a 20 mT/m readout gradient and 100 kHz receiver bandwidth.
(a)
Larmor frequency?
Write units the way you normally would: 1.5 MHz, 1500 kHz and 0.0015 GHz are the same answer.
(b)
Polarization in ppm?
(c)
Field of view?
Check
A spin-echo sequence with TR = 450 ms and TE = 20 ms, for a tissue with T1 = 2500 ms and T2 = 900 ms.
(a)
Relative signal?
(b)
Fraction of the longitudinal magnetization recovered by each pulse?
Check
A 0.6 kHz RF pulse with a 6 mT/m slice gradient; 500 MBq of ¹⁸F (half-life 109.8 min).
(a)
Slice thickness?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
(b)
Activity after 120 min?
Check
Heart muscle with rheobase 0.8 mA and chronaxie 0.3 ms, stimulated with 0.5 ms pulses into 500 Ω.
(a)
Threshold current?
Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.
(b)
Pulse energy at threshold?
(c)
Energy at the chronaxie instead?
Check
A pacemaker using 1 µC per pulse at 60 per minute, with 10 µA of circuit current, on a 2 Ah cell.
(a)
Average current?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
Battery life in years?
Check
A defibrillator: 30 µF at 1.5 kV into 75 Ω, truncated after 8 ms.
(a)
Stored energy?
Write units the way you normally would: 1.5 J, 1500 mJ and 0.0015 kJ are the same answer.
(b)
Peak current?
(c)
Energy delivered?
Check
A device with 4000 pF from its 120 V, 50 Hz supply to the case; ground resistance 1 Ω; a person offering 1 kΩ.
(a)
Leakage current?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
Current through the person with the ground intact?
Check
Grounds 8 mV apart across a 400 Ω catheter path; an isolated 230 V, 60 Hz system with 50 nF to ground.
(a)
Catheter current?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
Hazard current?
Check
A wearable with 3 mA active, 2 µA asleep, duty 0.02, on a 500 mAh cell.
(a)
Average current?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
Life in days?
Check
A 250 sample/s, 16-bit stream at 20 nJ/bit; tissue (σ = 1 S/m, ρ = 1000 kg/m³) in a 20 V/m field; an MTBF of 30 years.
(a)
Radio energy per day?
Write units the way you normally would: 1.5 J, 1500 mJ and 0.0015 kJ are the same answer.
(b)
SAR?
(c)
Probability of surviving 3 years?
Check
A dialyzer runs at 250 mL/min of blood flow, lowering urea from 90 to 40 mg/dL across it, for a 4-hour session on a 65 kg patient (body water 58% of weight).
(a)
Clearance, in mL/min?
Write units the way you normally would: 1.5 mL/min, 1500 µL/min and 0.0015 L/min are the same answer.
(b)
Kt/V?
(c)
Urea reduction ratio, ignoring generation?
Check
A screening test has sensitivity 0.9 and specificity 0.85, used where the prevalence is 0.1.
(a)
Positive predictive value?
(b)
Positive likelihood ratio?
(c)
Out of 1000 people screened, how many test positive?
Check
A study must estimate a sensitivity near 0.9 with a 95% confidence half-width of 0.03, in a condition of prevalence 0.1.
(a)
How many patients with the condition does it need?
(b)
How many people would an unselected study screen?
Have paper and a calculator ready. Work each problem out by hand, then enter the answer. Your first try on each one counts toward your score.
Practice
Keep practicing
15 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.