S15 · Lesson 31 · Respiratory measurements
Lung volumes and capacities, spirometry and the forced expiration with FEV1 over FVC, correcting gas volumes to body conditions, pneumotachographs, measuring the volume a spirometer can't reach by helium dilution and body plethysmography, and ventilation and end-tidal carbon dioxide
Breathing is easy to see and surprisingly hard to measure well. The volume a patient blows out depends on the temperature and humidity of the gas, so the same breath measures differently in the lungs and in the instrument. Some air can never be blown out at all, so measuring total lung capacity takes a gas-mixing trick or a sealed box and Boyle's law. And the single most used lung test, the forced expiration, reduces to a ratio of two volumes that says whether the airways are narrowed.
- A simple model of a forced expiration: the exhaled volume rises as , where the time constant τ is airway resistance times lung compliance.
- FEV1, the volume out in the first second, over the forced vital capacity: . Narrowed airways (asthma, COPD) raise τ and lower the ratio, below 0.7 by the usual criterion.
- Restrictive disease shrinks FVC but leaves the ratio normal or high.
Predict
A spirometer at 23 °C records 4.0 L exhaled. Corrected to body temperature (37 °C) and saturation with water vapor, is the true lung volume larger or smaller?
In the lungs the gas was warmer and fully saturated, so it occupied more volume than it does after cooling in the instrument. The BTPS factor is (310/296) × (760 − 21.1)/(760 − 47) = 1.085 at 23 °C and sea level, so the 4.0 L breath was 4.34 L in the lungs.
Name it
Volumes and capacities
A quiet breath moves the tidal volume (about 0.5 L). Beyond it lie the inspiratory and expiratory reserve volumes, and after a full exhalation the residual volume remains. Vital capacity is the most that can be exhaled after a full inhalation; total lung capacity adds the residual volume; functional residual capacity (FRC) is what's left at the end of a quiet breath, the expiratory reserve plus the residual volume. Spirometry can't measure the residual volume, so it can't give FRC or TLC.
Correcting gas volumes
Instruments measure gas at ambient temperature and pressure, saturated (ATPS). Body conditions (BTPS) are 37 °C and saturated, with water vapor at 47 mmHg:
Spirometers and pneumotachographs
Older spirometers collect gas in a bell; modern ones measure flow and integrate it. A pneumotachograph passes the flow through a small, linear resistance (a fine screen or capillary bundle) and measures the pressure drop, , with a differential pressure sensor (lessons 3 and 4). In a forced expiration the key results are FVC, FEV1, their ratio and the peak flow.
Volumes a spirometer can't reach
- Helium dilution: a spirometer of volume holding helium at fraction is connected at FRC; after mixing, , so .
- Body plethysmography: in a sealed box, the subject pants against a closed shutter; compressing the lung gas by ΔV raises its pressure by ΔP, and Boyle's law gives , with P the absolute pressure of the dry alveolar gas.
Ventilation and CO₂
Minute ventilation is tidal volume × rate; alveolar ventilation subtracts the anatomical dead space (about 150 mL) from each breath. A capnograph measures exhaled CO₂ by infrared absorption; end-tidal CO₂ of about 5% corresponds to ≈ 36 mmHg, close to arterial.
Worked example
Helium at 10% in a 5 L spirometer falls to 6.5% after the subject breathes in from FRC. What's the FRC? And what does a body box give for ΔV = 30 mL and ΔP = 10 cmH₂O at P = 970 cmH₂O?
Show the solution
- Helium dilution.
- Body box.
- Why they differ. The body box reads a little larger, as is typical, because it also counts gas trapped behind closed airways.
Answer helium dilution gives an FRC of 2.69 L, and the body box 2910 mL.
Worked example
In the sim's model with FVC = 4.5 L and τ = 0.4 s, find FEV1, the ratio and the peak flow. What's the alveolar ventilation at 500 mL tidal volume, 150 mL dead space and 14 breaths per minute?
Show the solution
- FEV1. a ratio of 0.918.
- Peak flow.
- Alveolar ventilation. out of a minute ventilation of 7 L/min.
Answer FEV1 is 4.13 L, a ratio of 0.918, the peak flow is 11.25 L/s (675 L/min), and the alveolar ventilation is 4.9 L/min.
Worked example
Arterial blood has a of 40 mmHg, and mixed expired gas 28 mmHg. What fraction of each 500 mL breath is dead space?
Show the solution
- Bohr's equation. It treats expired gas as alveolar gas diluted by dead-space gas with no CO₂:
- Dead space per breath. That's 150 mL of each 500 mL breath, which only moves in and out of the airways.
- Clinical use. A rising ratio, as in a pulmonary embolism that cuts perfusion to ventilated lung, shows up as a widening gap between arterial and expired CO₂.
Answer 0.3 of each breath, 150 mL, is dead space.
Check
A spirometer at 23 °C (water vapor pressure 17.5 mmHg) and barometric pressure 760 mmHg measures a volume. By what factor must it be multiplied to get the BTPS volume?
Check
A 5 L spirometer starts with 8% helium; after the subject joins at FRC and the gas mixes, it reads 6.5%. What's the FRC?
Write units the way you normally would: 1.5 L and 1500 mL are the same answer.
Check
In a body box, panting against a closed shutter compresses the lung gas by 20 mL and raises its pressure by 15 cmH₂O, from an absolute dry-gas pressure of 950 cmH₂O. What's the lung volume?
Write units the way you normally would: 1.5 mL, 1500 µL and 0.0015 L are the same answer.
Check
In the single-exponential model of a forced expiration with time constant 0.6 s, what's FEV1/FVC?
Check
A patient breathes 500 mL per breath at 20 breaths per minute with 120 mL of dead space. What's the alveolar ventilation?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
Check
End-tidal gas contains 5% CO₂ (of the dry gas) at a barometric pressure of 700 mmHg. What's its CO₂ partial pressure in the saturated alveolar gas?
Check
Arterial is 40 mmHg and mixed expired 22 mmHg, with a tidal volume of 600 mL.
(a)
What fraction of each breath is dead space?
(b)
How many milliliters is that?
Write units the way you normally would: 1.5 mL, 1500 µL and 0.0015 L are the same answer.
Check
A patient breathes V mL per breath at f breaths per minute, with D mL of dead space.
Step 1
What is the minute ventilation, in mL/min?
Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.
Step 2
What is the alveolar ventilation?
Explain why
Why can't a spirometer measure total lung capacity, and how do dilution and plethysmography get around that?
A spirometer measures only gas that moves in or out of the mouth, and even after the hardest possible exhalation the residual volume stays in the lungs, so every spirometer volume is a difference from some unknown starting amount. Helium dilution adds a known amount of an insoluble gas and measures how much it's diluted once it has mixed into the lungs, which reveals the total gas volume it spread into. Body plethysmography uses Boyle's law instead: compressing the gas in the chest slightly and measuring the pressure change gives the gas volume directly, including gas in regions that don't communicate well with the mouth, which dilution can miss.
Practice
Keep practicing
10 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.