S1532 of 56
0 XP
  1. Hook
  2. Play
  3. Predict
  4. Name it
  5. Check
  6. Explain

S15 · Lesson 32 · Respiratory mechanics and gas analysis

The lungs and chest wall as resistance and compliances, transpulmonary pressure, laminar pneumotachometers, the respiratory system's impedance by forced oscillation, infrared, paramagnetic and electrochemical gas analyzers, nitrogen washout and the diffusing capacity for carbon monoxide

Lesson 31 measured how much air the lungs hold and move. A patient with emphysema and one with lung fibrosis can have similar volumes, yet one has floppy lungs and narrowed airways while the other has stiff lungs and normal airways, and they need different treatment. Telling them apart takes the lungs' mechanical properties, resistance and compliance, measured the way an electrical engineer would measure a circuit: apply a pressure, measure a flow, and fit a model. And getting oxygen into the blood depends on gas transfer, which needs fast, accurate gas analyzers.

  • The respiratory system modeled as a series circuit: airway resistance R, the inertance I of the moving gas and tissue, and the total compliance C of lungs and chest wall. A loudspeaker superimposes small pressure oscillations on breathing (the forced oscillation technique), and the device measures the impedance Z=R+j(2πfI−1/(2πfC)).
  • The curves are the magnitude of Z and its reactance against frequency. Below resonance the compliance dominates and the reactance is negative; above it the inertance takes over.
  • Narrowed airways raise R across all frequencies. Stiffer lungs (smaller C) push the resonance up. Clinicians read both from one short measurement during quiet breathing, with no forced effort.

Predict

The lungs alone have a compliance of 0.2 L/cmH₂O and the chest wall also 0.2 L/cmH₂O. What's the compliance of the whole respiratory system?

Name it

The mechanical model

Webster models breathing with three pressures: at the airway opening, in the alveoli, and in the pleural space between lungs and chest wall. Between them sit:

  • airway resistance, Raw=(Pawo−PA)/Q, typically 1 to 2 cmH₂O·s/L in healthy adults;
  • lung compliance, CL=ΔV/Δ(Pawo−PPL) measured when nothing is flowing, about 0.2 L/cmH₂O;
  • chest-wall compliance, CW=ΔV/Δ(PPL−PBS) measured with the muscles relaxed, also about 0.2 L/cmH₂O.

The pleural pressure is estimated with a balloon in the esophagus. Lung and chest-wall compliances are in series, so the total is CLCW/(CL+CW). The product of resistance and compliance is the time constant of passive emptying. Fibrosis lowers compliance; emphysema raises compliance and resistance; asthma raises resistance.

Measuring flow

A pneumotachometer passes the flow through a small resistance and measures the pressure drop. A Fleisch type uses a bundle of fine parallel tubes so the flow stays laminar and, by Poiseuille's law, each tube drops

ΔP=8ηLπr4F

for a flow F through it (η the gas viscosity, about 1.8 × 10⁻⁵ Pa·s for air). The element is heated so exhaled water vapor doesn't condense, and its calibration changes with the gas's viscosity.

Impedance by forced oscillation

Superimposing small pressure oscillations at 5 to 35 Hz on quiet breathing and measuring the resulting flow gives the respiratory impedance. Its real part is resistance; its reactance is negative at low frequencies (compliance) and crosses zero at a resonant frequency of about 5 to 10 Hz in adults, f0=1/(2πIC), above which inertance dominates.

Gas analyzers

  • Infrared absorption measures CO₂ (strongly absorbed near 4.3 µm), N₂O and anesthetic agents by Beer's law, fast enough for breath-by-breath capnography.
  • Paramagnetic analyzers measure O₂, the only common respiratory gas strongly attracted by a magnetic field. Electrochemical (fuel cell) O₂ sensors are cheaper and slower.
  • Mass spectrometers separate every gas by mass at once; emission spectroscopy (an electric discharge that makes N₂ glow) measures nitrogen.

Gas transport

Nitrogen washout measures functional residual capacity like helium dilution does. The subject breathes pure oxygen from FRC, and all exhaled gas is collected. The nitrogen in it came from the lungs, so

FRC=VEFE,N2FA,N2,start−FA,N2,end

The diffusing capacity for carbon monoxide measures how readily gas crosses from alveoli into blood. In the single-breath method the subject inhales a mixture with a little CO, holds for about 10 s, and exhales; the CO's fall in the alveolar volume VA gives

DL,CO=VAt(PB−47)ln⁡FA,CO,0FA,CO,t

in mL/(min·mmHg), with 47 mmHg the water vapor pressure at body temperature. Normal adult values are about 25 to 30.

Worked example

A Fleisch pneumotachometer has 400 tubes, each 1 mm in diameter and 3 cm long. What pressure drop does 1 L/s of air produce, and what does that mean for the patient?

Show the solution
  1. Flow per tube. q=1400L/s=2.5×10−6m3/s
  2. Poiseuille's law. The drop is 8μLq/πr4, with 8μLq=8×1.8×10−5×0.03×2.5×10−6πr4=π×(0.5×10−3)4
  3. Pressure drop. ΔP=55Pa=0.56cmH2O
  4. Effect on the patient. The device's resistance is 0.56 cmH₂O·s/L, a third or less of a healthy person's airway resistance, so breathing through it barely changes the breathing being measured.
  5. Range. Doubling the flow doubles the drop as long as the flow stays laminar; at high flows turbulence makes the drop grow faster, which limits the useful range.

Answer 1 L/s produces a drop of 55 Pa (0.56 cmH₂O), a resistance small enough that it barely changes the patient's breathing.

Worked example

A patient breathes pure oxygen from FRC for 7 minutes; the 40 L of collected exhaled gas contains 4.8% nitrogen. Alveolar nitrogen started at 80% and ended near zero. What's the FRC? And what diffusing capacity does a 10 s breath-hold give with VA = 5500 mL and alveolar CO falling from 0.25% to 0.146%, at sea level?

Show the solution
  1. Nitrogen washed out. VN2=40×0.048=1.92L all of it from the lungs.
  2. FRC. The lungs started at 80% nitrogen: FRC=1.920.80=2.4L
  3. Diffusing capacity. With t = 10 s = 1/6 min and PB−47 = 713 mmHg: DL,CO=5500(1/6)×713ln⁡0.250.146=46.3×0.538=24.9mL/(min·mmHg) a normal value.
  4. What lowers it. Thickened alveolar walls in fibrosis or lost capillary bed in emphysema lower it.

Answer the FRC is 2.4 L, and the diffusing capacity is 24.9 mL/(min·mmHg), a normal value.

Worked example

A patient breathes 0.5 L at 15 breaths per minute, with a total compliance of 0.1 L/cmH₂O and an airway resistance of 2 cmH₂O·s/L; inspiration lasts 2 s at constant flow. How much work does each breath take, and what power?

Show the solution
  1. Elastic work. Stretching the elastic lung and chest stores We=12V2/C=12×0.25/0.1=1.25cmH2O·L
  2. Resistive work. Pushing 0.25 L/s through the airways for 2 s costs Wr=RV˙V=2×0.25×0.5=0.25cmH2O·L
  3. Total per breath. With 1 cmH₂O·L = 0.0981 J: W=1.5cmH2O·L=0.147J
  4. Power. At 15 breaths a minute the muscles deliver 0.037 W.
  5. Which part dominates. The elastic part dominates at rest; with narrowed airways or fast breathing the resistive part takes over.

Answer each breath takes 1.5 cmH₂O·L, 0.147 J, and the muscles deliver 0.037 W.

Check

Lung compliance is 0.15 L/cmH₂O and chest-wall compliance 0.1 L/cmH₂O. What's the total respiratory compliance, in L/cmH₂O?

Check

A Fleisch pneumotachometer has 400 tubes, each 0.8 mm in diameter and 5 cm long. What pressure drop does 2 L/s of air (η = 1.8 × 10⁻⁵ Pa·s) produce?

Write units the way you normally would: 1.5 Pa, 1500 mPa and 0.0015 kPa are the same answer.

Check

A patient's airway resistance is 2 cmH₂O·s/L and total compliance 0.05 L/cmH₂O. What's the time constant of passive exhalation?

Write units the way you normally would: 1.5 s and 1500 ms are the same answer.

Check

A respiratory system has inertance 0.01 cmH₂O·s²/L and compliance 0.05 L/cmH₂O. At what frequency does its forced-oscillation reactance cross zero?

Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.

Check

After a nitrogen washout the collected 40 L of exhaled gas contains 3.5% nitrogen. Alveolar nitrogen started at 80% and ended near zero. What's the FRC?

Write units the way you normally would: 1.5 L and 1500 mL are the same answer.

Check

A single-breath test at sea level has VA = 5500 mL, an 8 s breath-hold, and alveolar CO falling from 0.3% to 0.2%. What's DL,CO, in mL/(min·mmHg)?

Check

A breath of 0.4 L is drawn at constant flow over 2 s, with compliance 0.05 L/cmH₂O and resistance 8 cmH₂O·s/L, 12 times a minute.

(a)

How much work does each breath take?

Write units the way you normally would: 1.5 J, 1500 mJ and 0.0015 kJ are the same answer.

(b)

What power do the respiratory muscles deliver?

Check

A lung with resistance R (cmH₂O·s/L) and compliance C (L/cmH₂O) exhales passively.

Step 1

What is its time constant, in seconds?

Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.

Explain why

Why do lung and chest-wall compliances combine like capacitors in series, and why does that make the respiratory system stiffer than either part?

Practice

Keep practicing

10 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.