S15 · Lesson 7 · Temperature sensors
Thermistors with the B-parameter equation and their large negative coefficient, platinum resistance thermometers, thermocouples with cold-junction compensation, self-heating errors, and infrared thermometers that read the body's thermal radiation
Body temperature is the oldest clinical measurement and still one of the most useful, but the range that matters is narrow: a few degrees around 37 °C, read to a tenth of a degree. That favors sensors with a large, repeatable change over a small span. Thermistors change by about 4% per degree, which makes them the default in clinical probes and in the catheter tips used for thermodilution (lesson 26). Platinum resistors are the accurate reference, thermocouples the rugged choice, and infrared sensors read the eardrum or forehead without touching it.
- An NTC thermistor's resistance, , from 30 to 42 °C.
- The temperature coefficient is about −4%/°C near body temperature, ten times a platinum resistor's.
- Raise B and the curve steepens: more sensitivity, more nonlinearity.
Predict
A 10 kΩ (at 25 °C) thermistor with B = 3950 K is at body temperature, 37 °C. About what's its resistance?
= 10 kΩ × = 5.99 kΩ. An NTC thermistor's resistance falls as it warms, here by about 4.1% per degree: a 0.1 °C change is 25 Ω, easy to resolve.
Name it
Thermistors
Sintered metal-oxide thermistors have a negative temperature coefficient that follows, over a modest range,
with T in kelvins and B typically 3000 to 4500 K. Over wide ranges the Steinhart–Hart equation, , fits to a few millikelvins.
Resistance thermometers
Platinum RTDs are nearly linear: with = 100 Ω at 0 °C and α = 0.00385/°C for standard Pt100 elements. They're stable and interchangeable but change only 0.385 Ω per degree.
Thermocouples
Two dissimilar metals joined at two junctions produce a voltage proportional to the temperature difference between the junctions: , with a Seebeck coefficient S of about 40 µV/°C for copper–constantan (type T). The reference junction's temperature must be measured separately (cold-junction compensation).
Self-heating
A measuring current heats the sensor by , where δ is its dissipation constant (about 1 mW/°C in still air for a small bead, more in liquid). Keeping in the microwatts keeps the error below a few hundredths of a degree.
Infrared thermometers
A body at absolute temperature T radiates per unit area, peaking near , about 9.3 µm at body temperature. Ear and forehead thermometers focus that radiation onto a thermopile and compare it with the sensor's own temperature.
Worked example
Find the predict question's thermistor's resistance at 36 °C, its coefficient at 37 °C, and its self-heating with 100 µA flowing, if δ = 1 mW/°C.
Show the solution
- Resistance at 36 °C. so one degree changes it by 252 Ω.
- Coefficient at 37 °C.
- Self-heating power. At 37 °C, 100 µA in 5.99 kΩ dissipates
- Temperature rise. With δ = 1 mW/°C: a reading error worth correcting or avoiding with a smaller current.
Answer R = 6.241 kΩ at 36 °C (252 Ω per degree), α = −4.11%/°C at 37 °C, and 100 µA warms the bead by 0.060 °C.
Worked example
What does a Pt100 read at 37 °C, and what does a type T thermocouple give for a 12 °C difference at 43 µV/°C?
Show the solution
- Pt100 at 37 °C. only 0.385 Ω per degree.
- Thermocouple output. small enough that a microvolt-level amplifier and a known reference junction are essential.
Answer the Pt100 reads 114.245 Ω, and the thermocouple gives 516 µV.
Worked example
A 10 kΩ thermistor (at 25 °C, B = 3500 K) is linearized around 37 °C with a parallel resistor. What value puts the inflection of the combination at 37 °C?
Show the solution
- Thermistor at 37 °C.
- Condition for the inflection. The parallel combination's resistance-against-temperature curve has its inflection at when
- Parallel resistor.
- What it buys. The combination is then nearly linear for several degrees either side, at the cost of sensitivity: its slope is well under the bare thermistor's.
Answer a 4.44 kΩ resistor in parallel puts the inflection at 37 °C.
Check
A thermistor is 10 kΩ at 25 °C with B = 3400 K. What's its resistance at 40 °C?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
Check
What's the temperature coefficient, in %/°C, of a thermistor with B = 3950 K at 25 °C?
Check
What's the resistance of a platinum RTD with = 100 Ω (α = 0.00385/°C) at 20 °C?
Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.
Check
A thermocouple with Seebeck coefficient 43 µV/°C has its junctions 5 °C apart. What voltage does it produce?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A 3 kΩ thermistor carries 50 µA and has a dissipation constant of 5 mW/°C. How much does self-heating raise its temperature?
Check
At what wavelength does a body at 37 °C radiate most strongly?
Write units the way you normally would: 1.5 µm, 1500 nm and 0.0015 mm are the same answer.
Check
A thermistor is 10 kΩ at 25 °C with B = 3000 K.
(a)
What is its resistance at 40 °C?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
What parallel resistor puts the combination's inflection at 40 °C?
Check
A platinum RTD has resistance R₀ at 0 °C and coefficient α.
Step 1
What is its resistance at T °C?
Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.
Step 2
What temperature does a reading of X ohms mean?
Explain why
Why are thermistors preferred over platinum resistors for clinical temperature probes, and what do you give up?
Clinical temperatures span only a few degrees, so what matters is resolving small changes. A thermistor changes by about 4% per degree near body temperature, ten times a platinum resistor's 0.385%, so a simple divider and a modest converter resolve hundredths of a degree, and its high resistance makes lead resistance negligible. The price is nonlinearity, since resistance is exponential in inverse temperature, and less interchangeability and long-term stability than platinum, which is why thermistor probes are calibrated or matched to curves while platinum remains the reference standard.