Power budgets with duty cycling, battery life, the energy cost of sending data versus processing it on the device, storage needs, specific absorption rate limits for radio and wireless power, and reliability with a constant failure rate

An ECG patch that records for two weeks, a continuous glucose monitor, a hearing aid, an implanted neurostimulator: all are embedded systems (C12) whose hardest constraint is energy. A coin cell holds a couple of kilojoules; a radio can burn through it in days if it streams raw data. Designing these devices means budgeting every microampere, choosing what to compute on the device and what to send, keeping radio and wireless-power exposure within safety limits, and making something that won't fail inside a person.

  • Battery life of a device that is active (sampling, processing, transmitting) a fraction D of the time and asleep otherwise, from a 220 mAh coin cell.
  • The average current is IaD+Is(1−D). At low duty cycles the sleep current dominates; at high ones the active current does.
  • Halving the active time doubles life only while active current dominates.

Predict

A sensor draws 5 mA when active and 5 µA asleep, and is active 1% of the time. About how long does a 220 mAh coin cell last?

Name it

Power budgets

With duty cycle D, active current Ia and sleep current Is:

Iavg=IaD+Is(1−D),life=capacityIavg

Real capacity is derated for temperature, pulse loads and the cutoff voltage. A coin cell of 220 mAh at 3 V stores about 2.4 kJ.

Compute or transmit

Radios cost roughly nanojoules per bit, while a microcontroller can process a sample for far less. Streaming a 250 sample/s, 16-bit ECG all day is 345.6 million bits, about 3.5 J/day at 10 nJ/bit; sending only beat times, rhythm alerts or compressed data cuts the radio's share by orders of magnitude. Storing raw data instead needs channels × rate × bytes × time of memory.

Exposure limits

Radio links and wireless power transfer heat tissue. The specific absorption rate,

SAR=σ|E|2ρ

(with E the RMS field, σ the tissue conductivity and ρ its density), is limited, for example to 1.6 W/kg averaged over 1 g of tissue in the United States and 2 W/kg over 10 g in Europe. Implants must also keep surface temperature rises to a degree or two.

Reliability

With a constant failure rate λ (or mean time between failures 1/λ), the probability of surviving a time t is R(t)=e−λt, and a system that fails if any of its n independent parts fails has a failure rate equal to the sum of theirs. Implants are designed with redundancy, hermetic packaging and conservative derating because replacement means surgery.

Worked example

Size the storage and radio energy for a one-day, single-lead ECG at 250 samples/s and 16 bits.

Show the solution
  1. Storage per day. N=250×16×86400=345.6Mbit=43.2MB
  2. Radio energy. Streaming it at 10 nJ/bit costs E=3.46J per day
  3. Against the battery. That's about 0.15% of a 2.4 kJ coin cell per day, which adds up to a big share over two weeks, so many patches store data locally and upload later.

Answer a day of ECG takes 43.2 MB of storage, and streaming it costs 3.46 J of radio energy.

Worked example

What SAR does a field of 30 V/m (RMS) produce in tissue with σ = 0.5 S/m and ρ = 1000 kg/m³? What's the chance a device with a 50-year MTBF survives 5 years?

Show the solution
  1. SAR. SAR=0.5×3021000=0.45W/kg within the 1.6 W/kg limit.
  2. Survival over 5 years. R=e−5/50=0.905 One device in ten fails within five years, which would be far too many for an implant.

Answer the SAR is 0.45 W/kg, and the device survives 5 years with a probability of 0.905.

Worked example

A wrist device harvests body heat with a thermoelectric module giving 25 µW/cm² at a 5 K skin-to-air difference. With 3 cm² of module and a 60% efficient boost converter, can it run a monitor that averages 30 µW?

Show the solution
  1. Module output. P=25μW/cm2×3cm2=75μW
  2. After the converter. P=0.6×75μW=45μW 15 µW more than the monitor uses on average, enough to keep a small battery topped up while the wearer is still.
  3. In a warm room. The temperature difference may halve, and thermoelectric power goes as its square, so the harvest falls to a quarter, 11 µW; the battery must then carry the deficit.
  4. Planning around it. Harvesters set a budget that the duty cycle must respect.

Answer yes, while the wearer is still: 45 µW reaches the load, 15 µW more than the monitor's 30 µW. In a warm room the harvest falls to 11 µW and the battery must carry the deficit.

Check

A device draws 5 mA when active and 1 µA asleep, active a fraction 0.1 of the time. What's its average current?

Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.

Check

A 220 mAh battery supplies an average of 20 µA. How many days does it last?

Check

A device streams 500 samples/s of 12-bit data all day by radio at 50 nJ/bit. How much energy does the radio use per day?

Write units the way you normally would: 1.5 J, 1500 mJ and 0.0015 kJ are the same answer.

Check

Tissue with conductivity 0.7 S/m and density 1050 kg/m³ sees an RMS field of 10 V/m. What's the specific absorption rate?

Write units the way you normally would: 1.5 W/kg, 1500 mW/kg and 0.0015 kW/kg are the same answer.

Check

A device has a constant failure rate with an MTBF of 50 years. What's the probability it survives 2 years?

Check

A device has three parts, any of which failing ends its life, with failure rates of 200, 100 and 80 per billion hours. What's its MTBF in years?

Check

A thermoelectric harvester gives 25 µW/cm² over 2 cm² through a converter of efficiency 0.7, powering a 60 µW average load.

(a)

What surplus (positive) or deficit (negative) power results?

Write units the way you normally would: 1.5 µW, 1500 nW and 0.0015 mW are the same answer.

(b)

What if the temperature difference falls to 0.5 of its value?

Check

A device draws a mA when active and s µA asleep, active a fraction D of the time, from a C mAh battery.

Step 1

What is its average current, in µA?

Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.

Explain why

Why do wearable devices process data on board instead of streaming everything to a phone?

Practice

Keep practicing

10 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.