Producing X-rays in a tube, exponential attenuation with half-value layers, contrast between bone and soft tissue, projections as line integrals, reconstructing slices by filtered back-projection, Hounsfield units, and the photon noise that sets dose

An X-ray image is a shadow: photons that make it through the body expose the detector, and bone, which absorbs more, leaves a light silhouette. A plain radiograph flattens the whole body onto one plane. Computed tomography measures thousands of such shadows from every angle around the patient and solves for the attenuation at every point in a slice, the inverse problem that won Hounsfield and Cormack the 1979 Nobel Prize. The mathematics is the Fourier-slice theorem of S7's world, and the limit on image quality is counting statistics: each pixel is only as good as the number of photons behind it.

  • The fraction of X-rays transmitted through soft tissue (μs, about 0.19 cm⁻¹ at diagnostic energies) of thickness L, with and without a 3 cm bone (μb) in the path.
  • Attenuation is exponential, so each half-value layer, ln⁡2/μ, halves the beam.
  • The contrast between paths with and without bone, 1−e−(μb−μs)b, doesn't depend on how much soft tissue surrounds it.

Predict

Diagnostic X-rays are attenuated by soft tissue with μ ≈ 0.19 cm⁻¹. What fraction passes straight through a 20 cm abdomen?

Name it

Making X-rays

Electrons accelerated through the tube voltage (60 to 140 kV) strike a tungsten anode. Most of their energy becomes heat; under 1% becomes bremsstrahlung and characteristic X-rays, with photon energies up to e times the tube voltage, so a 100 kV tube's shortest wavelength is hc/E = 1.24 nm·keV/100 keV = 0.0124 nm.

Attenuation

A narrow beam through a uniform material falls as

I=I0e−μx,HVL=ln⁡2μ

μ depends on photon energy and on the material's density and atomic number (photoelectric absorption grows roughly as Z3/E3), which is why calcium-rich bone and iodine contrast agents stand out. Through a path of different materials the exponents add: I=I0exp⁡(−∑μixi).

Projections and reconstruction

Each CT measurement gives a line integral of attenuation:

p=ln⁡I0I=∫μdl

The set of all projections at all angles is the Radon transform of the slice. Filtered back-projection filters each projection with a ramp filter |k| and smears it back across the image along its direction; summed over angles, that inverts the transform. Sampling needs about π2N views for N detector bins across the field.

Hounsfield units

CT reports attenuation relative to water:

HU=1000μ−μwμw−μair≈1000μ−μwμw

so water is 0, air −1000, soft tissues tens to a few tens of HU, and dense bone around +1000 or more.

Noise and dose

Photons arrive randomly, so a measurement built on N detected photons has relative noise 1/N. Halving the noise takes four times the photons, and four times the dose.

Worked example

A beam crosses 20 cm of soft tissue, or 17 cm of soft tissue plus 3 cm of bone with μ = 0.5 cm⁻¹. What's the contrast between the two paths, and what's one projection value?

Show the solution
  1. Contrast between the paths. The bone path transmits IbIt=e−(0.5−0.19)×3=0.395 times as much, a contrast of 60%.
  2. Soft tissue alone. II0=0.0224
  3. Projection value. p=ln⁡10.0224=3.80 which is just μL = 0.19 × 20.

Answer the bone path transmits 0.395 times as much, a 60% contrast, and the soft-tissue path's projection value is 3.80.

Worked example

A tissue has μ = 0.21 cm⁻¹ where water's is 0.19. What's its CT number? How many photons give 1% noise, and how many views does a 512-bin detector need?

Show the solution
  1. CT number. HU=1000×0.020.19=+105
  2. Photons for 1% noise. N=10.012=10000 per measurement.
  3. Views. A 512-bin detector needs about π2×512≈800 views around the patient.

Answer the tissue's CT number is +105, 1% noise takes 10,000 photons per measurement, and the detector needs about 800 views.

Worked example

A CT protocol gives acceptable noise at 2 mSv with 1 mm pixels and 5 mm slices. What dose would keep the same noise with 0.5 mm pixels and 2.5 mm slices?

Show the solution
  1. How noise scales. In CT the noise variance in a voxel goes as σ2∝1Dw3h with D the dose, w the pixel width and h the slice thickness.
  2. Why. Fewer photons pass through each smaller voxel, and the reconstruction filter amplifies the finer detail's noise.
  3. Dose for the smaller voxels. Halving w costs a factor of 8, and halving h another 2: D=2mSv×8×2=32mSv a sixteenfold rise.
  4. Consequence. That's why high-resolution scans are limited to small regions or accept more noise.

Answer it would take 32 mSv, sixteen times the dose.

Check

What fraction of a narrow X-ray beam passes through 5 cm of tissue with μ = 0.19 cm⁻¹?

Check

What's the half-value layer of a material with μ = 0.19 cm⁻¹?

Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.

Check

A voxel has μ = 0.23 cm⁻¹ and water's is 0.19 cm⁻¹. What's its CT number, taking air as zero?

Check

A CT detector sees 0.02 of the unattenuated intensity. What's the projection value, ∫μ dl?

Check

How many detected photons give a relative noise of 0.01?

Check

Bone with μ = 0.45 cm⁻¹ replaces 5 cm of soft tissue with μ = 0.19 cm⁻¹ in a beam's path. What's the contrast, the fractional drop in transmitted intensity?

Check

A CT scan at 1 mSv is changed: pixel width scaled by 0.5, slice thickness by 0.4, and target noise (standard deviation) by a factor of 0.7.

(a)

What dose does the new protocol need?

Check

A material has attenuation coefficient μ (cm⁻¹); water's is μw.

Step 1

What fraction of a narrow beam passes through x cm?

Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.

Explain why

Why does a CT image need projections from many angles, when a radiograph uses only one?

Practice

Keep practicing

10 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.