S15 · Lesson 15 · Biopotential amplifiers

What a biopotential amplifier must do, the three-op-amp instrumentation amplifier and its gain, common-mode rejection ratio and where it comes from, how mismatched electrode impedances turn common-mode voltage into a differential error, and splitting the gain around a high-pass filter

An ECG amplifier must find a 1 mV signal between two electrodes while both sit, together, at tens of millivolts or more of power-line voltage picked up by the patient's body, and each electrode adds its own few hundred millivolts of DC offset. It has to amplify the difference a thousandfold while rejecting the common part almost completely, without loading electrodes whose impedance can reach hundreds of kilohms. The instrumentation amplifier of C6 is the core of the answer, but the electrodes themselves set a limit that no amplifier can fix on its own.

  • The input-referred error from a common-mode voltage vcm: the amplifier's own rejection contributes vcm/CMRR, and a mismatch ΔZ between the two electrode impedances contributes vcmΔZ/Zcm, where Zcm is the input's common-mode impedance.
  • With a 100 dB amplifier, the electrodes usually dominate: 20 kΩ of mismatch against 100 MΩ is a CMRR of only 74 dB.
  • Raising the input impedance, or lowering the common-mode voltage itself (lesson 17), are the remedies.

Predict

The body sits at 10 mV of 60 Hz common-mode voltage. The two ECG electrodes differ in impedance by 20 kΩ, and the amplifier's common-mode input impedance is 100 MΩ. How large a 60 Hz differential error appears at its input?

Name it

Requirements

A biopotential amplifier needs differential gain of about 1000 for ECG (more for EEG), an input impedance far above the electrodes' (above 10 MΩ, ideally gigaohms in common mode), a high common-mode rejection ratio, a bandwidth matched to the signal, low noise, protection against defibrillation pulses, and isolation of the patient (lesson 24).

The instrumentation amplifier

Two noninverting input buffers sharing a gain resistor R1 (with feedback resistors R2), followed by a difference amplifier with gain R4/R3:

G=(1+2R2R1)R4R3

The first stage amplifies the differential signal but passes common-mode at unity gain, so the second stage's rejection is multiplied by the first stage's gain.

Common-mode rejection

With differential gain Gd and common-mode gain Gcm, CMRR=Gd/Gcm, usually quoted as 20log10⁡ of that. Referred to the input, a common-mode voltage appears as vcm/CMRR. A single difference amplifier with gain G and resistors of tolerance t has a worst-case CMRR of about (1+G)/(4t).

Electrode mismatch

Source impedances Z1 and Z2 feeding inputs of common-mode impedance Zcm convert common-mode voltage into a differential one:

vd≈vcmZ1−Z2Zcm

so the system's CMRR can't exceed Zcm/ΔZ, however good the amplifier.

Gain staging

Electrode offsets of up to ±300 mV limit the DC-coupled first stage to a gain of about 10 to 20 (lesson 14). A high-pass filter (0.05 Hz for diagnostic ECG, 0.5 Hz for monitoring) then removes the offset, and later stages supply the rest of the gain.

Worked example

An instrumentation amplifier has R1 = 5.6 kΩ, R2 = 25 kΩ and a unity-gain difference stage. What's its gain? How much more gain does a 1000-times ECG front end need after the high-pass filter?

Show the solution
  1. Gain of the instrumentation amplifier. G=1+2×255.6=9.93 about 9.9 before the filter, keeping a 300 mV offset under 3 V.
  2. Gain still needed. A total of 1000 needs 10009.93=101 more after the filter.

Answer the gain is 9.93, and the front end needs 101 times more after the high-pass filter.

Worked example

A difference amplifier with gain 10 is built from 0.1% resistors. What CMRR can it guarantee, and what's the total input-referred 60 Hz error with 10 mV common-mode, a 100 dB amplifier and the predict question's electrodes?

Show the solution
  1. CMRR from 0.1% resistors. CMRR≈1+104×0.001=2750 or 68.8 dB, which is why integrated instrumentation amplifiers with laser-trimmed resistors are used.
  2. The amplifier's share. With 100 dB of CMRR: eamp=10mV105=0.1μV
  3. The electrodes' share. The electrode mismatch adds 2 µV.
  4. Total. About 2.1 µV, dominated by the electrodes.

Answer the resistors can guarantee a CMRR of 2750 (68.8 dB), and the total input-referred error is about 2.1 µV, dominated by the electrodes.

Worked example

An ECG amplifier has 10 nV/√Hz of input voltage noise and 1 pA/√Hz of input current noise, and each input sees 50 kΩ of electrode. What's the total input noise over 150 Hz?

Show the solution
  1. Current noise through the electrode. ei=1pA/Hz×50kΩ=50nV/Hz
  2. The other two sources. The electrode's own thermal noise is 4kTR = 29 nV/√Hz, and the amplifier adds 10 nV/√Hz.
  3. Add them in power. They're independent: en=102+502+292=59nV/Hz
  4. Over 150 Hz. vn=59×150nV=0.72μV rms
  5. What sets the floor. With high-impedance electrodes the amplifier's current noise, not its voltage noise, sets the floor, so dry-electrode front ends use FET inputs with femtoampere noise currents.

Answer the total input noise is 59 nV/√Hz, 0.72 µV RMS over 150 Hz.

Check

An instrumentation amplifier has gain resistor R1 = 5.6 kΩ, feedback resistors R2 = 10 kΩ and a difference stage of gain 10. What's its gain?

Check

An amplifier has differential gain 100 and common-mode gain 0.001. What's its CMRR in decibels?

Check

A 20 mV common-mode voltage reaches an amplifier with 80 dB of CMRR. How large an input-referred error does that alone produce?

Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.

Check

A 20 mV common-mode voltage reaches inputs of 1000 MΩ common-mode impedance through electrodes that differ by 5 kΩ. What differential voltage does the mismatch create?

Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.

Check

A difference amplifier with gain 5 uses resistors of tolerance 0.0001. What's its worst-case CMRR in dB?

Check

A 300 mV electrode offset must not push a first stage beyond ±3 V. If that stage uses all the gain it can, how much gain must follow the high-pass filter for a total of 2000?

Check

An amplifier's input noise is 4 nV/√Hz and 0.01 pA/√Hz; each input sees 1000 kΩ of source resistance at 37 °C, over a 150 Hz band.

(a)

What is the total noise density?

Write units the way you normally would: 1.5 nV/√Hz, 1500 pV/√Hz and 0.0015 µV/√Hz are the same answer.

(b)

What RMS noise is that over the band?

Check

A three-op-amp instrumentation amplifier has gain resistor R1, feedback resistors R2 and a difference stage of gain k.

Step 1

What gain does its input stage have for differential signals?

Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.

Explain why

Why can't a perfect amplifier reject common-mode interference if the electrode impedances differ?

Practice

Keep practicing

10 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.