S15 Biomedical instrumentation · Practice
Biopotential amplifiers
Problems dealt one at a time, with new numbers every time. The kinds you miss come back more often, and you can keep going as long as you like.
Check
An instrumentation amplifier has gain resistor = 5.6 kΩ, feedback resistors = 10 kΩ and a difference stage of gain 10. What's its gain?
Check
An amplifier has differential gain 100 and common-mode gain 0.001. What's its CMRR in decibels?
Check
A 20 mV common-mode voltage reaches an amplifier with 80 dB of CMRR. How large an input-referred error does that alone produce?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A 20 mV common-mode voltage reaches inputs of 1000 MΩ common-mode impedance through electrodes that differ by 5 kΩ. What differential voltage does the mismatch create?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A difference amplifier with gain 5 uses resistors of tolerance 0.0001. What's its worst-case CMRR in dB?
Check
A 300 mV electrode offset must not push a first stage beyond ±3 V. If that stage uses all the gain it can, how much gain must follow the high-pass filter for a total of 2000?
Check
An amplifier's input noise is 4 nV/√Hz and 0.01 pA/√Hz; each input sees 1000 kΩ of source resistance at 37 °C, over a 150 Hz band.
(a)
What is the total noise density?
Write units the way you normally would: 1.5 nV/√Hz, 1500 pV/√Hz and 0.0015 µV/√Hz are the same answer.
(b)
What RMS noise is that over the band?
Check
INA gain: = 0.5 kΩ, = 5 kΩ, second stage 10?
Check
Mismatch error: 1 mV, ΔZ = 1 kΩ, = 5000 MΩ?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
Difference-amp CMRR: G = 1, tolerance 5 × 10⁻⁵?