S15 · Lesson 49 · Biotelemetry
Sending physiological signals by radio, sampling and bit rates for telemetered biopotentials, the frequency bands for medical telemetry and implants, how tissue absorbs radio waves, the energy cost per bit, and powering implants through the skin by inductive coupling
Webster calls radiotelemetry the best isolation there is: a battery-powered transmitter on the patient has no connection at all to the mains. Hospital cardiac monitors have long used it, and today every implant that talks to a programmer or a phone, and every one recharged through the skin, depends on a wireless link through tissue that absorbs radio waves and limits how much power may be deposited in it. The design questions are how many bits must be sent, at what frequency, with what energy, and how much power can be pushed in.
- Two coils couple a fraction k of their magnetic flux, and each has quality factor Q. The curve is the best power-transfer efficiency the link can reach with an optimal load, , against k.
- Implants sit centimeters from their external coils, so k is a few percent. High-Q coils rescue the link: what matters is the product kQ, not k alone.
- Efficiency rises steeply once kQ passes about 1.
Predict
An implant's coil couples k = 0.05 to an external coil; both have Q = 100. What's the best power-transfer efficiency?
With = 0.0025 × 10⁴ = 25, the optimum efficiency is 25/(1 + √26)² = 0.67. Resonating both coils lets energy slosh between them many times per cycle, so a weak coupling transfers power efficiently as long as the coils lose little per cycle. That is why implant chargers use resonant, high-Q coils.
Name it
Telemetry systems
A telemetry transmitter amplifies the signal, digitizes it and modulates a carrier; the receiver recovers the data. Webster's rule is to sample at least five times the highest frequency of interest, though twice suffices in theory, so the bit rate is
Telemetry is immune to ground loops but not to interference at its own frequency, nor to dropouts when the antennas' orientations null the link, so good systems report signal loss.
Frequency bands
Hospital patient telemetry in the US uses the Wireless Medical Telemetry Service bands (608 to 614, 1395 to 1400 and 1427 to 1432 MHz) to avoid interference from other users. Implants use the Medical Implant Communication Service band, 402 to 405 MHz, with channels up to 300 kHz wide and transmitted power limited to 25 µW EIRP, and wearables commonly use Bluetooth Low Energy at 2.4 GHz.
Radio through tissue
Tissue is a lossy dielectric: muscle at 403 MHz has a relative permittivity of about 57 and a conductivity of about 0.8 S/m. A plane wave's field decays as with
about 19 Np/m for muscle at 403 MHz: 1.65 dB per centimeter. Higher frequencies lose more, which is one reason implants use the 400 MHz band rather than 2.4 GHz. The absorbed power is also limited by the SAR rules of lesson 48.
Energy per bit
A radio's energy cost is roughly constant per bit sent, from nanojoules for low-power links to tens of nanojoules with overhead, so average power is the bit rate times the energy per bit, and sending features or detected events rather than raw samples saves energy in proportion.
Inductive power
Implants that need more power than a primary battery can supply for their lifetime are powered or recharged by a coil outside the body coupled to one inside. For coupling coefficient k and quality factors and , the best efficiency, with the load chosen optimally, is
Coupling falls steeply with distance and misalignment, so links are designed for the worst position. The implant can send data back by changing its coil's load, which the external reader sees as a change in its own coil's current (load modulation), with no transmitter of its own.
Worked example
A three-lead ECG telemeter keeps components up to 150 Hz, samples at five times that with 12 bits, and transmits at 5 nJ per bit. What bit rate and average radio power does it need?
Show the solution
- Sample rate. Five times 150 Hz is 750 samples a second per lead.
- Bit rate.
- Radio power. At 5 nJ per bit: That's 3.24 mWh a day, 0.9 mAh from a 3.6 V cell.
- Sending less. Sending only detected beats and their intervals, a few hundred bits a minute, would cut that by a factor of a thousand.
Answer the telemeter needs 27 kb/s, and its radio averages 135 µW.
Worked example
An implant needs 10 mW delivered to its electronics. Its coil couples k = 0.02 to the external coil, with Q = 100 outside and 50 inside. What's the best efficiency, and how much power must the external unit put into its coil?
Show the solution
- Coupling figure.
- Best efficiency.
- External power. and the remaining 27 mW heats the coils and tissue.
- Closer coils. If k doubles, rises to 8 and the efficiency to 0.50, so the external coil needs 20 mW instead of 37.
Answer the best efficiency is 0.27, so the external unit must put 37 mW into its coil.
Worked example
How much does muscle attenuate a 403 MHz signal from an implant 3 cm deep, and how does that compare with 2.45 GHz, where muscle has a relative permittivity of about 53 and a conductivity of about 1.7 S/m?
Show the solution
- At 403 MHz. With α ≈ 19 Np/m, 3 cm costs
- At 2.45 GHz. The same formula gives α ≈ 45 Np/m, 3.9 dB/cm, so the path costs 11.7 dB, more than twice as much.
- Near-field coupling. The shorter wavelength also makes the antenna's near-field coupling to tissue worse.
- Which band. The lower band wins for deep implants, though the higher one offers more bandwidth for shallow wearables.
Answer 3 cm of muscle costs 5.0 dB at 403 MHz and 11.7 dB at 2.45 GHz, more than twice as much.
Check
A telemeter sends 3 channels with content up to 40 Hz, sampled at five times that with 16 bits, at 50 nJ per bit.
(a)
What bit rate does it need?
Write units the way you normally would: 1.5 kb/s, 1500 b/s and 0.0015 Mb/s are the same answer.
(b)
What average radio power does that take?
Check
An implant's coil couples k = 0.03 to an external coil; their quality factors are 50 and 80, and the implant needs 50 mW.
(a)
What is the best efficiency the link can reach?
(b)
How much power must the external coil deliver?
Write units the way you normally would: 1.5 mW, 1500 µW and 0.0015 W are the same answer.
Check
Tissue attenuates a radio wave's field by 34 Np/m.
(a)
How many decibels does a path of 1 cm cost?
(b)
At what depth has the field fallen to 1/e?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
Check
Tissue has relative permittivity 57 and conductivity 0.5 S/m, at 2450 MHz.
(a)
What is its loss tangent, σ/ωε?
(b)
What is its field attenuation constant?
Check
An implant's link has a 100 kHz channel and needs 50 kb/s.
(a)
What spectral efficiency does it need?
Write units the way you normally would: 1.5 b/s/Hz and 0.0015 kb/s/Hz are the same answer.
Check
Why do implant data links favor the 400 MHz band over 2.4 GHz?
Check
Two resonant coils with coupling k and quality factors and transfer power to an optimal load.
Step 1
What figure of merit sets the link's efficiency?
Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.
Step 2
What is the best efficiency?
Explain why
How can two coils that share only a few percent of their magnetic flux transfer most of the power between them?
In a resonant link, each coil is tuned with a capacitor to the operating frequency, so energy oscillates between its magnetic and electric fields many times, Q times per radian, before it is lost. The receiving coil captures only a small fraction k of the transmitting coil's flux on each cycle, but because neither coil loses much energy per cycle, the exchange repeats until the receiving side's load has taken most of it. What matters is how fast energy crosses compared with how fast it is lost, which is the product kQ: once it exceeds about one, the efficiency is high even though k itself is small.
Practice
Keep practicing
8 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.