A telemeter sends 3 channels with content up to 40 Hz, sampled at five times that with 16 bits, at 50 nJ per bit.
(a)
What bit rate does it need?
Write units the way you normally would: 1.5 kb/s, 1500 b/s and 0.0015 Mb/s are the same answer.
(b)
What average radio power does that take?
Check
An implant's coil couples k = 0.03 to an external coil; their quality factors are 50 and 80, and the implant needs 50 mW.
(a)
What is the best efficiency the link can reach?
(b)
How much power must the external coil deliver?
Write units the way you normally would: 1.5 mW, 1500 µW and 0.0015 W are the same answer.
Check
Tissue attenuates a radio wave's field by 34 Np/m.
(a)
How many decibels does a path of 1 cm cost?
(b)
At what depth has the field fallen to 1/e?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
Check
Tissue has relative permittivity 57 and conductivity 0.5 S/m, at 2450 MHz.
(a)
What is its loss tangent, σ/ωε?
(b)
What is its field attenuation constant?
Check
An implant's link has a 100 kHz channel and needs 50 kb/s.
(a)
What spectral efficiency does it need?
Write units the way you normally would: 1.5 b/s/Hz and 0.0015 kb/s/Hz are the same answer.
Check
Bit rate: 2 channels to 100 Hz, 14 bits, 5× sampling?
Write units the way you normally would: 1.5 kb/s, 1500 b/s and 0.0015 Mb/s are the same answer.
Check
Link efficiency: k = 0.02, Q = 80 and 100?
Check
Attenuation: 20 Np/m over 2 cm?
A kind of problem counts as solid once you've got it right on the first try twice in a row. Back to the lesson · Practice every topic in S15