S15 · Lesson 35 · Clinical laboratory instruments
Spectrophotometry with Beer's law and the absorbance that minimizes error, calibration against standards, automated analyzers, chromatography with internal standards, electrophoresis, and the hematology analyzer's impedance counting, red-cell indices and coincidence
Most of the numbers in a patient's chart never touch the patient's body directly. A tube of blood goes to the laboratory, where analyzers run hundreds of tests an hour: glucose and cholesterol by the color a reagent turns, drug levels by how long a compound takes to leave a column, and blood counts by pushing cells one at a time through a hole 80 µm across and watching the resistance jump. Each is an electrical measurement of something chemical or cellular. How does a spectrophotometer turn transmitted light into a concentration, and why does it work best when the sample absorbs about two-thirds of it?
- A spectrophotometer reads transmittance T and reports concentration through the absorbance , which is proportional to concentration.
- Every photometer has some fixed error in reading T. The curve is the resulting relative error in concentration, in percent, against the sample's absorbance, for the transmittance error (in percent of full scale) set by the slider.
- The error is smallest at A = 0.434 (T = 37%) and grows quickly at both ends. That's why labs dilute samples that read above A ≈ 1 and choose reactions that give A well above 0.1.
Predict
A sample transmits 10% of the light. A second sample of the same substance, at twice the concentration in the same cuvette, transmits what fraction?
Beer's law makes absorbance proportional to concentration: 10% transmittance is A = 1, so twice the concentration gives A = 2, which is 1% transmittance. Each layer of absorber removes the same fraction of what reaches it, so transmittance falls exponentially with concentration and path length.
Name it
Beer's law
Light of power passing a path L through a solution of concentration C leaves with
where a is the absorptivity of the substance at that wavelength. Transmittance is reported as a percentage, so . A spectrophotometer has a source, a wavelength selector (filter or monochromator) set to the absorption peak of the reaction's colored product, a cuvette, a photodetector and readout.
Calibration
With a and L fixed, concentration follows from a standard of known concentration : . That one-point method assumes Beer's law holds across the range. Solvent absorption, stray light and cuvette reflections bend the line, so labs run several standards spanning the range and read unknowns from the calibration curve.
Where the photometer is most accurate
A fixed error ΔT in reading transmittance gives a concentration error
which is smallest at . Webster's working range is about 0.11 to 1.0; outside it, errors grow quickly.
Automated analyzers, chromatography and electrophoresis
Chemistry analyzers automate the same photometry: they pipette sample and reagent into cuvettes on a rotor or track, time the reaction and read absorbance, often at two wavelengths to cancel turbidity, at hundreds to thousands of tests an hour.
A chromatograph separates a mixture as a carrier gas or liquid sweeps it along a column, each compound held back by its affinity for the stationary phase. Compounds are identified by retention time and quantified by peak area. A known amount of an internal standard added to every sample corrects for losses: the unknown's concentration is the standard's times the ratio of peak areas, divided by a response factor from calibration. Two peaks are separated when their resolution reaches about 1.5.
Electrophoresis separates charged molecules, such as serum proteins, by moving them through a gel or paper in an electric field E. Each travels a distance set by its mobility μ. Heating from the current raises mobility, so the voltage or current is held constant.
The hematology analyzer
Blood is diluted in a conducting solution and drawn through an aperture between two electrodes, about 50 to 100 µm wide. The aperture's resistance is , and each cell displaces conducting fluid, raising the resistance by about for a cell of volume v (about 1.5 times that for a sphere, from how the current bends around it). With a constant current, each cell makes a voltage pulse whose count gives the cell count and whose height gives the cell's volume. Counts are corrected for coincidence, two cells in the aperture at once, which statistics predicts from the concentration and sensing volume. Red cells, white cells and platelets are counted at different dilutions; white cells are also classified by light scatter in a flow cell (flow cytometry).
From the red-cell count (millions per µL), hematocrit (percent) and hemoglobin (g/dL), the red-cell indices follow:
Normal ranges are about 82 to 98 fL, 27 to 31 pg and 32 to 36%.
Worked example
A standard of 10 mg/dL reads 40% transmittance and a patient sample 55%. What's the patient's concentration, and is the measurement in a good range?
Show the solution
- Absorbances.
- Concentration.
- Is the range good? Both absorbances are inside the 0.11 to 1.0 range, near the 0.434 optimum, so a 0.5% transmittance error costs only about 1.4% (standard) to 1.5% (sample) in concentration.
- Linearity. The one-point calculation is only as good as Beer's law over this range; a lab would confirm linearity with standards above and below.
Answer the patient's concentration is 6.52 mg/dL, measured in a good range: both absorbances sit near the 0.434 optimum.
Worked example
A hematology aperture is 80 µm across and 100 µm long, filled with diluent of resistivity 1.5 Ω·m, and driven by 1 mA. How big is the pulse from a 90 fL red cell?
Show the solution
- Aperture area.
- Aperture resistance. so 1 mA puts 29.8 V across it.
- Change from one cell. A 90 fL cell (90 µm³) raises the resistance by about
- Pulse size. That's a 5.3 mV pulse on 29.8 V, a change of 1.8 parts in 10,000: the fraction of the aperture's volume the cell fills.
- What the analyzer does with it. It counts thousands of such pulses a second, and a histogram of their heights gives the volume distribution, from which the MCV and the spread of sizes (RDW) are computed.
Answer a 90 fL red cell gives a pulse of about 5.3 mV on the 29.8 V across the aperture.
Worked example
Blood with 5 million red cells per microliter is diluted 1:50,000 and drawn through an 80 µm, 100 µm long aperture. How often is a second cell in the aperture when one is counted?
Show the solution
- Diluted concentration. Diluting 5 million cells per microliter 1:50,000 leaves 100 cells per µL.
- Aperture volume.
- Mean number inside.
- Chance of a second cell. By Poisson statistics, the chance that another cell is present along with the one being counted is
- Consequence. Such coincidences merge two pulses into one, undercounting and inflating apparent volume, so analyzers apply a coincidence correction that grows with concentration.
Answer a second cell is in the aperture 4.9% of the time a cell is counted.
Check
A sample reads 25% transmittance. What's its absorbance?
Check
A standard of 8 mg/dL reads 20% transmittance; the unknown reads 60%. What's the unknown's concentration, in mg/dL?
Check
A photometer reads transmittance to within ±0.2% (of full scale). What's the relative concentration error for a sample with A = 0.434?
Check
A blood count gives hematocrit 42% and red cells 3.8 million per µL. What's the MCV, in fL?
Check
Hemoglobin is 13.5 g/dL and hematocrit 34%. What's the MCHC?
Check
An impedance-counting aperture is 50 µm in diameter and 100 µm long, filled with diluent of resistivity 1.2 Ω·m. What's its resistance?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
Check
A protein with mobility 5 × 10⁻⁵ cm²/(V·s) migrates in a field of 5 V/cm. How far does it move in 60 minutes?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
Check
Blood with 5 million cells per µL is diluted 1:10000 and drawn through a 100 µm diameter aperture 100 µm long.
(a)
How many cells does the aperture hold on average?
(b)
What is the chance another cell is present when one passes?
Check
A blood count gives hematocrit H%, red cells R million per µL and hemoglobin b g/dL.
Step 1
What is the mean cell volume, in femtoliters?
Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.
Step 2
What is the mean cell hemoglobin, in picograms?
Step 3
What is the mean cell hemoglobin concentration, in g/dL?
Explain why
Why does a spectrophotometer measure concentration least accurately when the sample absorbs almost nothing or almost everything?
Concentration is proportional to absorbance, the logarithm of the transmitted fraction, while the photometer's errors are roughly fixed amounts of transmittance. When the sample absorbs almost nothing, T is near 1 and the absorbance is a tiny number computed from the small difference between two nearly equal readings, so a fixed error in T is a large fraction of it. When the sample absorbs almost everything, T is tiny, and a fixed error in T is a large fraction of T itself, which the logarithm turns into a large error in absorbance. In between the two effects balance: the relative error is smallest at A = 0.434, where 37% of the light gets through. Labs dilute strong samples and choose reactions and path lengths that put absorbances in the middle of the range.