S15 Biomedical instrumentation · Practice
Clinical laboratory instruments
Problems dealt one at a time, with new numbers every time. The kinds you miss come back more often, and you can keep going as long as you like.
0 solved
0 in a row
0 of 11 kinds solid
Check
A sample reads 25% transmittance. What's its absorbance?
Check
A standard of 8 mg/dL reads 20% transmittance; the unknown reads 60%. What's the unknown's concentration, in mg/dL?
Check
A photometer reads transmittance to within ±0.2% (of full scale). What's the relative concentration error for a sample with A = 0.434?
Check
A blood count gives hematocrit 42% and red cells 3.8 million per µL. What's the MCV, in fL?
Check
Hemoglobin is 13.5 g/dL and hematocrit 34%. What's the MCHC?
Check
An impedance-counting aperture is 50 µm in diameter and 100 µm long, filled with diluent of resistivity 1.2 Ω·m. What's its resistance?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
Check
A protein with mobility 5 × 10⁻⁵ cm²/(V·s) migrates in a field of 5 V/cm. How far does it move in 60 minutes?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
Check
Blood with 5 million cells per µL is diluted 1:10000 and drawn through a 100 µm diameter aperture 100 µm long.
(a)
How many cells does the aperture hold on average?
(b)
What is the chance another cell is present when one passes?
Check
Absorbance at 8% transmittance?
Check
MCV in fL for HCT 39% and RBC 3.4 million/µL?
Check
Standard 4.1 mg/dL at 17%T, unknown at 57%T. Concentration in mg/dL?
A kind of problem counts as solid once you've got it right on the first try twice in a row. Back to the lesson · Practice every topic in S15