S15 Biomedical instrumentation · Practice
Diagnostic accuracy, device software and regulation
Problems dealt one at a time, with new numbers every time. The kinds you miss come back more often, and you can keep going as long as you like.
0 solved
0 in a row
0 of 9 kinds solid
Check
Against a reference standard, a test is positive in 90 of the patients who have the disease and negative in 3 of them. What's its sensitivity?
Check
Among healthy subjects the test is negative in 900 and positive in 20. What's its specificity?
Check
A test with sensitivity 0.95 and specificity 0.9 is used where the prevalence is 0.2. What's its positive predictive value?
Check
For the same test and prevalence (sensitivity 0.95, specificity 0.9, prevalence 0.2), what's the negative predictive value?
Check
What's the positive likelihood ratio of a test with sensitivity 0.9 and specificity 0.9?
Check
An arrhythmia detector falsely flags 2 beats in 100,000. How many false alarms does it raise for a patient with a heart rate of 60 beats/min monitored for 72 hours?
Check
A study must show a sensitivity of about 0.95 with a 95% confidence half-width of 0.02, in a condition with prevalence 0.1.
(a)
How many patients with the condition must it include?
(b)
How many people would an unselected screening study need?
Check
PPV: sensitivity 0.74, specificity 0.64, prevalence 0.11?
Check
LR+ for sensitivity 0.68, specificity 0.55?
A kind of problem counts as solid once you've got it right on the first try twice in a row. Back to the lesson · Practice every topic in S15