S15 · Lesson 23 · Direct blood-pressure measurement
Measuring arterial pressure through a fluid-filled catheter, the catheter and transducer as a second-order system with natural frequency and damping set by their dimensions, why a small air bubble ruins the response, the fast-flush test, and hydrostatic offsets
In intensive care and in the catheter lab, arterial pressure is measured beat by beat through a thin plastic tube filled with saline, connected to a disposable silicon bridge transducer (lesson 3). The column of liquid has mass, the transducer's diaphragm and the tubing have compliance, and the narrow tube has viscous friction: a mass, a spring and a damper. Whether the trace on the monitor is the patient's real pressure waveform or a ringing caricature of it depends on the numbers lesson 2 introduced, and this lesson shows where those numbers come from.
- A catheter of radius 0.5 mm and length 1 m, filled with saline, connected to a transducer whose compliance is 0.01 mm³ per 100 mmHg. A small air bubble of volume adds compliance .
- The amplitude ratio of the whole system against frequency. Without the bubble, is 163 Hz and the response is flat across the pressure waveform's harmonics.
- A 1 mm³ bubble drops to 43 Hz: the harmonics near 20 Hz get exaggerated, though the damping rises too.
Predict
A 1 mm³ air bubble gets into a catheter system whose transducer has a compliance of 0.01 mm³ per 100 mmHg. What happens to its natural frequency?
Air at atmospheric pressure compresses by per pascal: 1 mm³/101 kPa = 9.9 × 10⁻¹⁵ m³/Pa, thirteen times the transducer's 7.5 × 10⁻¹⁶. Natural frequency goes as the square root of stiffness, so it falls by √14.2 ≈ 3.8, from 163 to 43 Hz.
Name it
The catheter as a second-order system
Model the liquid in a catheter of radius r and length L as a slug of mass , the transducer's diaphragm as a spring of volume stiffness , and Poiseuille friction in the tube as a damper. Then
with ρ and η the liquid's density and viscosity (about 1000 kg/m³ and 0.001 Pa·s for saline). Wider, shorter catheters and stiffer transducers raise ; damping falls steeply with radius.
Bubbles
Compliances add, and a gas bubble is very compliant: isothermally, . Even 1 mm³ dominates a modern transducer. A bubble lowers (as ) and raises ζ (as ); the frequency effect usually matters more, because the waveform's harmonics then fall near the resonance.
The fast-flush test
Opening the flush valve briefly and releasing it applies a pressure step, and the system rings. If successive oscillation peaks shrink by a ratio d, then
where is the ringing frequency, one over the time between peaks.
Units and leveling
1 mmHg = 133.32 Pa. The transducer reads pressure relative to its own height, so it's leveled with the heart: a height error h adds , about 0.74 mmHg per centimeter of saline.
Worked example
Find and ζ for the sim's catheter without a bubble, and with the 1 mm³ bubble.
Show the solution
- Transducer compliance. so = 1.33 × 10¹⁵ Pa/m³.
- Natural frequency without a bubble.
- Damping without a bubble.
- With the bubble. The compliance becomes 1.06 × 10⁻¹⁴ m³/Pa, which gives
Answer without the bubble, = 163 Hz and ζ = 0.016; with the 1 mm³ bubble, = 43 Hz and ζ = 0.059.
Worked example
A flush test shows successive peaks shrinking to 0.3 of the previous one, 50 ms apart. Find ζ and . And what error does a transducer 10 cm below the heart give?
Show the solution
- Log of the decay.
- Damping ratio.
- Natural frequency. Peaks 50 ms apart ring at 20 Hz: an underdamped, low-frequency system that will overstate systolic pressure.
- Level error. A 10 cm level error adds a saline column of (7.8 mmHg if the column were blood).
Answer ζ = 0.188 and = 20.4 Hz, and a transducer 10 cm below the heart adds 7.4 mmHg to the reading.
Worked example
A transducer is placed 5 cm below a patient's right atrium. What error does that cause in a central venous pressure of 6 mmHg, and in an arterial pressure of 100 mmHg?
Show the solution
- Saline column. added to every reading.
- Arterial pressure. On 100 mmHg that's 3.7%.
- Central venous pressure. On 6 mmHg it's 61%: the reading would say 9.7 mmHg and suggest fluid overload where there is none.
- The remedy. That's why the transducer is leveled to the phlebostatic axis, at the fourth intercostal space halfway through the chest, and re-zeroed whenever the patient's bed moves.
Answer the transducer adds 3.7 mmHg to every reading: 3.7% of an arterial pressure of 100 mmHg, but 61% of a central venous pressure of 6 mmHg, which would read 9.7 mmHg.
Check
A saline-filled catheter of radius 0.5 mm and length 0.5 m feeds a transducer whose compliance is 0.05 mm³ per 100 mmHg. What's the natural frequency?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
For that catheter (radius 0.5 mm, length 0.5 m, transducer compliance 0.05 mm³ per 100 mmHg, η = 0.001 Pa·s), what's the damping ratio?
Check
A catheter system has = 150 Hz with a transducer compliance of 7.5 × 10⁻¹⁶ m³/Pa. A 3 mm³ air bubble enters. What's the new natural frequency?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
In a fast-flush test, each oscillation peak is 0.25 of the one before. What's the damping ratio?
Check
A pressure transducer sits 10 cm below the heart, connected by a liquid column of density 1000 kg/m³. How much does it overread?
Check
What's 120 mmHg in kilopascals?
Write units the way you normally would: 1.5 kPa, 1500 Pa and 0.0015 MPa are the same answer.
Check
A pressure wave with heart rate 90 per minute is recorded through a catheter system with = 15 Hz and ζ = 0.6.
(a)
What is the frequency of its tenth harmonic?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
By what factor does the system scale that harmonic's amplitude?
Check
A transducer sits a height h (m) below the reference level, connected by liquid of density ρ (kg/m³), while measuring a pressure P (mmHg).
Step 1
What offset does the liquid column add, in pascals?
Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.
Step 2
What percent error is that?
Explain why
Why does a tiny air bubble in a pressure line distort the arterial waveform so much?
The catheter system resonates where the mass of the liquid column bounces on the elasticity of everything that can change volume: the transducer's diaphragm, the tubing walls and any gas. A modern transducer is extremely stiff, so even a cubic millimeter of air, which compresses easily, can be ten times more compliant than everything else combined. The added compliance lowers the natural frequency in proportion to the square root of the stiffness, moving the resonance down into the range of the pressure waveform's harmonics, where they're amplified and the trace overshoots and rings. That's why lines are carefully flushed of bubbles and checked with a flush test.