S15 Biomedical instrumentation · Practice
Direct blood-pressure measurement
Problems dealt one at a time, with new numbers every time. The kinds you miss come back more often, and you can keep going as long as you like.
Check
A saline-filled catheter of radius 0.5 mm and length 0.5 m feeds a transducer whose compliance is 0.05 mm³ per 100 mmHg. What's the natural frequency?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
For that catheter (radius 0.5 mm, length 0.5 m, transducer compliance 0.05 mm³ per 100 mmHg, η = 0.001 Pa·s), what's the damping ratio?
Check
A catheter system has = 150 Hz with a transducer compliance of 7.5 × 10⁻¹⁶ m³/Pa. A 3 mm³ air bubble enters. What's the new natural frequency?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
In a fast-flush test, each oscillation peak is 0.25 of the one before. What's the damping ratio?
Check
A pressure transducer sits 10 cm below the heart, connected by a liquid column of density 1000 kg/m³. How much does it overread?
Check
What's 120 mmHg in kilopascals?
Write units the way you normally would: 1.5 kPa, 1500 Pa and 0.0015 MPa are the same answer.
Check
A pressure wave with heart rate 90 per minute is recorded through a catheter system with = 15 Hz and ζ = 0.6.
(a)
What is the frequency of its tenth harmonic?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
By what factor does the system scale that harmonic's amplitude?
Check
: r = 0.2 mm, L = 0.2 m, compliance 0.1 mm³/100 mmHg?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
ζ from a peak ratio of 0.2?
Check
Level error: 1 cm, ρ = 1000?
A kind of problem counts as solid once you've got it right on the first try twice in a row. Back to the lesson · Practice every topic in S15