S1546 of 56
0 XP
  1. Hook
  2. Play
  3. Predict
  4. Name it
  5. Check
  6. Explain

S15 · Lesson 46 · Electrical safety of medical equipment

What current does to the body, macroshock and microshock thresholds, skin and body resistance, leakage current through stray capacitance, why the ground wire matters, ground-potential differences and catheters, isolated power systems and line isolation monitors, and patient isolation

In a hospital, patients are connected to electrical equipment through low-resistance paths that ordinary appliances never provide: gel electrodes that bypass dry skin, and catheters filled with conductive saline that lead straight to the heart. A current too small to feel through the hands can trigger fibrillation when it enters the heart directly. Electrical safety in medicine is therefore an engineering discipline with its own numbers: how much current is dangerous by which path, how much leaks through every power supply's stray capacitance, and what grounding, isolation and monitoring keep it below the limits.

  • The current through a person who touches 120 V, 60 Hz, against their body resistance, compared with rough thresholds for current entering through the skin: perception (about 0.5 mA), let-go (about 10 mA) and ventricular fibrillation (about 100 mA).
  • Dry skin keeps an accidental touch near the perception level; wet or broken skin, or an electrode with gel, lets the same voltage drive a dangerous current.
  • For current applied directly to the heart, the danger level drops by about a factor of a thousand.

Predict

A device's power supply has 2500 pF of stray capacitance between the 120 V, 60 Hz line and its metal case. How much current flows through someone touching the case if its ground wire is broken?

Name it

Physiological effects

For 60 Hz current entering and leaving through the skin of the limbs (macroshock), typical thresholds are: perception about 0.5 mA; let-go (the largest current at which a person can still release a grip) about 6 to 10 mA; ventricular fibrillation roughly 75 to 400 mA; burns and sustained contraction at amperes. Thresholds rise at frequencies above about 1 kHz, which is why electrosurgery can pass large currents at several hundred kilohertz. When current enters the heart directly through a catheter or pacing lead (microshock), fibrillation can occur at tens to hundreds of microamperes, so patient connections that reach the heart are limited to about 10 µA.

Resistance of the body

Dry, intact skin is the main barrier, from tens to hundreds of kilohms; wet or broken skin, or gel electrodes, drop it to around a kilohm. The body's interior is a few hundred ohms. Current is set by I=V/R along whatever path exists.

Leakage current and grounding

Every line-powered device leaks current to its case through stray and filter capacitance: I=V2πfC. The protective ground wire carries it away; the current divides between the ground path and a person touching the case in inverse proportion to their resistances, so an intact ground (a fraction of an ohm) takes nearly all of it. A broken ground sends it all through the person. Limits for chassis leakage are about 100 µA in normal conditions.

Ground-potential differences

Two grounded devices can sit at slightly different potentials because ground currents flow through the wiring's resistance. A patient connected to both, through a saline catheter of a few hundred ohms, carries ΔV/R: a few millivolts can exceed 10 µA. Equipotential grounding, bonding all equipment near the patient to one point, prevents this.

Isolation

Isolated power systems in operating rooms supply equipment from an isolation transformer, so a single fault to ground doesn't complete a circuit; a line isolation monitor continuously measures the current that would flow through a fault, the hazard current V2πfC of the system's capacitance to ground, and alarms at 5 mA. Isolated patient inputs (lesson 15's front ends behind isolation barriers) keep patient leads floating so they can't become a return path.

Worked example

What current does a 120 V touch drive through dry skin (100 kΩ) and through a wet or gelled path (1 kΩ)?

Show the solution
  1. Through dry skin. I=120V100kΩ=1.2mA just above perception.
  2. Through a wet or gelled path. I=120V1kΩ=120mA in the fibrillation range for a hand-to-hand path.
  3. What decides the outcome. The voltage is the same; the path decides the outcome.

Answer the touch drives 1.2 mA through dry skin and 120 mA through the wet path.

Worked example

Two outlets near a patient differ by 20 mV. A saline catheter and the patient's body form a 500 Ω path between equipment on each. What flows, and what potential difference would keep it under 10 µA? What hazard current does a line isolation monitor report for 30 nF of system capacitance at 120 V, 60 Hz?

Show the solution
  1. Current through the catheter path. I=20mV500Ω=40μA four times the microshock limit.
  2. Allowed difference. Staying under 10 µA requires less than V=10μA×500Ω=5mV between grounds.
  3. Line isolation monitor. It reports I=120×2π×60×30nF=1.36mA under its 5 mA alarm limit.

Answer 40 µA flows, the grounds must differ by less than 5 mV to keep it under 10 µA, and the monitor reports 1.36 mA.

Worked example

A hot wire touches a grounded metal case. The fault path has 0.2 Ω of wiring and 0.1 Ω in the ground conductor. What fault current flows, what voltage does the case reach before the breaker opens, and what flows through a person (1 kΩ) touching it?

Show the solution
  1. Fault current. On a 120 V circuit: I=120V0.3Ω=400A far above a 20 A breaker's rating, so it trips in a few milliseconds.
  2. Case voltage. Until it does, the case sits above ground at V=400×0.1=40V
  3. Current through a person. 40 V drives 40 mA through 1 kΩ: dangerous, but for milliseconds.
  4. Without the ground conductor. The case would sit at 120 V indefinitely and the breaker would never trip; the ground wire's job is to turn a shock hazard into a short circuit.

Answer 400 A flows, the case reaches 40 V until the breaker opens, and 40 mA flows through the person for those few milliseconds.

Check

A person touches 120 V through a total path resistance of 1 kΩ. What current flows?

Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.

Check

A device has 10000 pF between its 120 V, 50 Hz supply and its case. What leakage current flows through that capacitance?

Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.

Check

Two pieces of equipment connected to a patient differ in ground potential by 20 mV, and the path between them through a catheter is 300 Ω. What current flows?

Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.

Check

300 µA of leakage reaches a case connected to ground through 0.1 Ω. A person touching it offers a 10 kΩ path to ground. How much flows through the person?

Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.

Check

An isolated power system at 120 V, 50 Hz has 80 nF of capacitance to ground. What hazard current does its line isolation monitor report?

Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.

Check

A catheter path has 300 Ω, and the microshock limit is 10 µA. What's the largest ground-potential difference allowed between the two connected devices?

Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.

Check

A hot-to-case fault occurs on a 120 V circuit with 0.1 Ω of wiring and 0.5 Ω in the ground conductor; a person of 5 kΩ touches the case.

(a)

What fault current flows?

Write units the way you normally would: 1.5 A, 1500 mA and 0.0015 kA are the same answer.

(b)

What current flows through the person until the breaker trips?

Check

A device has C pF between its V-volt, f-hertz supply and its case, which is grounded through g ohms; a person of R ohms touches the case.

Step 1

What leakage current flows, in microamperes?

Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.

Explain why

Why is a current that's harmless through the hands dangerous through a catheter?

Practice

Keep practicing

10 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.