S15 Biomedical instrumentation · Practice
Electrical safety of medical equipment
Problems dealt one at a time, with new numbers every time. The kinds you miss come back more often, and you can keep going as long as you like.
Check
A person touches 120 V through a total path resistance of 1 kΩ. What current flows?
Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.
Check
A device has 10000 pF between its 120 V, 50 Hz supply and its case. What leakage current flows through that capacitance?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
Two pieces of equipment connected to a patient differ in ground potential by 20 mV, and the path between them through a catheter is 300 Ω. What current flows?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
300 µA of leakage reaches a case connected to ground through 0.1 Ω. A person touching it offers a 10 kΩ path to ground. How much flows through the person?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
An isolated power system at 120 V, 50 Hz has 80 nF of capacitance to ground. What hazard current does its line isolation monitor report?
Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.
Check
A catheter path has 300 Ω, and the microshock limit is 10 µA. What's the largest ground-potential difference allowed between the two connected devices?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
A hot-to-case fault occurs on a 120 V circuit with 0.1 Ω of wiring and 0.5 Ω in the ground conductor; a person of 5 kΩ touches the case.
(a)
What fault current flows?
Write units the way you normally would: 1.5 A, 1500 mA and 0.0015 kA are the same answer.
(b)
What current flows through the person until the breaker trips?
Check
Body current: 5 V, 0.5 kΩ?
Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.
Check
Leakage: 100 V, 50 Hz, 20000 pF?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
Microshock: 1 mV over 100 Ω?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
A kind of problem counts as solid once you've got it right on the first try twice in a row. Back to the lesson · Practice every topic in S15