S15 · Lesson 17 · Interference and the driven right leg
How the power line couples into a patient and the leads through stray capacitance and magnetic fields, the common-mode voltage on the body, differential pickup on unshielded leads, shielding and twisting, and the driven-right-leg circuit that actively cancels the common-mode voltage
In a hospital room a patient is surrounded by power wiring, and a few picofarads of stray capacitance are enough to drive a fraction of a microampere of 60 Hz current through the body. That current has to reach ground somewhere, and on its way it raises the whole body by millivolts, a common-mode voltage the amplifier must reject. Current that leaks into an unshielded lead wire does worse: it flows through one electrode's impedance and arrives as a genuine differential signal. This lesson sizes each path, and then shows the trick that every modern ECG front end uses: drive the patient's body with an inverted copy of its own common-mode voltage.
- The body's common-mode voltage with a driven-right-leg circuit: , where is the displacement current from the power line, the protective resistor to the right-leg electrode and the loop's gain.
- With no loop gain the body simply sits at ; feedback divides that by the loop gain plus one.
- Compare it with a plain 50 kΩ ground electrode, the dashed line.
Predict
A patient is coupled to a 120 V, 60 Hz line through 3 pF of stray capacitance and grounded through a 50 kΩ electrode. What common-mode voltage does the body sit at?
The stray capacitance has a reactance of 1/(2π × 60 × 3 pF) = 884 MΩ, so it drives = 120 V/884 MΩ = 0.136 µA, a pure displacement current. Flowing out through 50 kΩ, it raises the body by 6.8 mV. The amplifier then needs about 77 dB of rejection to bring that below a microvolt.
Name it
Capacitive coupling to the body
The power line drives a displacement current through the stray capacitance C between it and the patient. That current returns to ground through the ground (right-leg) electrode, impedance , raising the body's common-mode voltage to . The amplifier's CMRR and the electrodes' mismatch (lesson 15) decide how much of it becomes differential error.
Coupling to the leads
Displacement current coupled into a lead wire flows to the body through that lead's electrode. If the two electrodes' impedances differ, the currents produce a differential voltage directly at the amplifier's inputs. With nanoamperes of current and tens of kilohms of mismatch this reaches hundreds of microvolts, which is why ECG leads are shielded.
Magnetic coupling
A changing magnetic field through the loop formed by the leads and the body induces . Keeping the leads together, twisted, shrinks the area A.
The driven right leg
Instead of grounding the right leg, an amplifier senses the common-mode voltage (the average of the two input buffers' outputs, through resistors ), inverts and amplifies it (feedback resistor ), and drives the right leg through a protective resistor . The body's common-mode voltage becomes
The patient sees the right leg through a large resistor, so a fault can't drive much current through them, while the loop gain makes the effective ground impedance small.
Worked example
A driven-right-leg circuit has = 5 MΩ, = 25 kΩ and = 5 MΩ, with = 0.2 µA. What's the common-mode voltage, and what effective ground impedance does the patient see?
Show the solution
- Common-mode voltage. against 10 mV for a 50 kΩ ground electrode.
- Effective ground impedance. lower than the electrode.
- Fault protection. A fault current is still limited by the full 5 MΩ.
Answer the common-mode voltage is 2.49 mV, and the patient sees an effective ground impedance of 12.5 kΩ.
Worked example
9 nA couples into one lead, and the electrodes differ by 20 kΩ. What differential pickup results? And what does a 1 µT, 60 Hz field induce in a 0.05 m² lead loop?
Show the solution
- Differential pickup. a fifth of the ECG itself, and not rejected by any CMRR.
- Its cure. Shielding the leads removes it.
- Magnetic pickup. The field induces cut by twisting the leads together.
Answer the mismatch gives 180 µV of differential pickup, and the magnetic field induces 18.8 µV in the loop.
Worked example
A battery-powered ECG monitor floats: its circuit common has 200 pF to earth, and the patient couples 2 pF to a 120 V, 60 Hz line. What voltage do patient and monitor take relative to earth, what current flows through the patient, and what common-mode voltage reaches the amplifier through a 50 kΩ reference electrode?
Show the solution
- Series capacitance. The 2 pF and 200 pF in series give
- Current through the patient.
- Floating voltage. Patient and monitor float together at from earth, but that voltage is common to the inputs and the circuit common alike, so the amplifier never sees it.
- Common-mode voltage at the amplifier. What it sees is the current's drop across the reference electrode: which 80 dB of CMRR turns into 0.45 µV at the input.
- Why floating helps. Floating the whole amplifier with the patient is why battery-powered and isolated front ends reject line interference so well.
Answer patient and monitor float at 1.19 V from earth, 89.6 nA flows through the patient, and 4.5 mV of common-mode voltage reaches the amplifier.
Check
A 120 V, 50 Hz line couples to a patient through 10 pF. What displacement current flows?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
A displacement current of 0.2 µA leaves a patient through a 10 kΩ ground electrode. What's the body's common-mode voltage?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
9 nA of line current couples into one ECG lead, and the electrode impedances differ by 5 kΩ. How large a differential pickup results?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A 60 Hz magnetic field of 1 µT threads a 0.01 m² loop formed by the leads and the body. What voltage is induced?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A driven-right-leg circuit has = 5 MΩ, = 5 MΩ and = 100 kΩ, with 0.1 µA of displacement current. What's the body's common-mode voltage?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
A 10 mV common-mode voltage must contribute less than 0.5 µV of input-referred error. What CMRR is needed?
Check
A floating monitor's common has 1000 pF to earth, and the patient couples 1 pF to a 120 V, 60 Hz line. The reference electrode is 100 kΩ, and the amplifier's CMRR is 60 dB.
(a)
What current flows through the patient?
Write units the way you normally would: 1.5 nA, 1500 pA and 0.0015 µA are the same answer.
(b)
What common-mode voltage reaches the amplifier?
(c)
What input-referred error results?
Check
A driven-right-leg circuit has output resistor , feedback resistor and averaging resistors (MΩ), with displacement current (µA).
Step 1
What effective resistance does it present to the body, in MΩ?
Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.
Step 2
What common-mode voltage results, in volts?
Explain why
How does the driven-right-leg circuit reduce interference and also make the patient safer?
Power-line current coupled into the body must leave through some path, and the voltage it develops across that path is the common-mode voltage the amplifier must reject. A driven-right-leg circuit measures that voltage, inverts and amplifies it, and drives the body through the right-leg electrode in the opposite sense, so negative feedback holds the body close to the amplifier's reference: the common-mode voltage is divided by one plus the loop gain, as if the patient were connected to ground through a very small impedance. Because the drive passes through a large protective resistor, the patient isn't actually connected to ground through a low impedance, so a fault elsewhere can't drive a dangerous current through them via the right-leg electrode.
Practice
Keep practicing
10 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.