S15 Biomedical instrumentation · Practice
Interference and the driven right leg
Problems dealt one at a time, with new numbers every time. The kinds you miss come back more often, and you can keep going as long as you like.
Check
A 120 V, 50 Hz line couples to a patient through 10 pF. What displacement current flows?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
A displacement current of 0.2 µA leaves a patient through a 10 kΩ ground electrode. What's the body's common-mode voltage?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
9 nA of line current couples into one ECG lead, and the electrode impedances differ by 5 kΩ. How large a differential pickup results?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A 60 Hz magnetic field of 1 µT threads a 0.01 m² loop formed by the leads and the body. What voltage is induced?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A driven-right-leg circuit has = 5 MΩ, = 5 MΩ and = 100 kΩ, with 0.1 µA of displacement current. What's the body's common-mode voltage?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
A 10 mV common-mode voltage must contribute less than 0.5 µV of input-referred error. What CMRR is needed?
Check
A floating monitor's common has 1000 pF to earth, and the patient couples 1 pF to a 120 V, 60 Hz line. The reference electrode is 100 kΩ, and the amplifier's CMRR is 60 dB.
(a)
What current flows through the patient?
Write units the way you normally would: 1.5 nA, 1500 pA and 0.0015 µA are the same answer.
(b)
What common-mode voltage reaches the amplifier?
(c)
What input-referred error results?
Check
Displacement current: 100 V, 50 Hz, 20 pF?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
DRL: = 0.5 MΩ, = 0.05 µA, = 10 MΩ, = 200 kΩ?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Magnetic pickup: 0.1 µT, 0.005 m², 60 Hz?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
A kind of problem counts as solid once you've got it right on the first try twice in a row. Back to the lesson · Practice every topic in S15