S15 · Lesson 1 · Medical measurement systems

The chain from physiological measurand to display, what makes medical measurements hard, sensitivity and full-scale output, nonlinearity measured against the endpoint line, interfering and modifying inputs, resolution referred to the input, and combining independent errors

Medical instruments measure quantities that are small, slow, buried in other signals and attached to a person who moves, sweats and must never be harmed. An ECG is about a millivolt sitting on top of a few hundred millivolts of electrode offset and a volt of power-line pickup. A blood-pressure transducer has to read a few millimeters of mercury accurately while its temperature drifts. This course works through how each kind of medical measurement is made, from sensor to amplifier to processed number, using the circuits, signals and physics of the earlier courses. It starts with the vocabulary every instrument specification uses.

  • A pressure sensor whose output bends slightly away from a straight line: y=Sx+ax2 over a range of 0 to X mmHg.
  • The endpoint line joins the outputs at zero and full scale. The largest gap between the curve and that line, as a fraction of the full-scale output, is the nonlinearity.
  • For a quadratic bend the gap peaks at mid-range, at |a|X2/4.

Predict

A blood-pressure transducer has a sensitivity of 5 µV per volt of excitation per mmHg and runs from 5 V. What does it put out at 100 mmHg?

Name it

The generalized instrument

A measurand (pressure, potential, flow, temperature, concentration) reaches a sensor, which converts it to an electrical quantity. Signal conditioning amplifies, filters and digitizes it, and the result is displayed, stored or used to control therapy. Calibration signals and feedback may be part of the loop. Measurements are direct (a catheter in an artery), indirect (a cuff on the arm) or by sampling (blood drawn for the lab).

Inputs that aren't the measurand

  • Interfering inputs add to the output: power-line pickup, electrode offset, a temperature-driven zero drift.
  • Modifying inputs change the sensitivity itself: temperature changing a strain gauge's gain, supply voltage changing a bridge's output.

Remedies are negative feedback, filtering, opposing inputs (a second, matched sensor that sees only the interference), and shielding.

Static characteristics

  • Sensitivity: output change per unit input. Bridge sensors are quoted per volt of excitation.
  • Accuracy: closeness to the true value; precision: repeatability; resolution: the smallest change that can be detected.
  • Nonlinearity: the largest deviation from a reference line, as a percent of full scale. Against the endpoint line, a quadratic term ax2 deviates by at most |a|X2/4.
  • Resolution referred to the input: an N-bit converter spanning a range R after a gain G resolves R/(2NG) at the input.

Independent error sources combine as the root of the sum of squares.

Worked example

A transducer gives 25 µV/mmHg with a bend of −0.002 µV/mmHg² over 0 to 300 mmHg. Find its full-scale output and its nonlinearity.

Show the solution
  1. Full-scale output. At 300 mmHg: V=25×300−0.002×3002=7500−180=7320μV
  2. Largest gap from the endpoint line. It falls at 150 mmHg: Δ=0.002×30024=45μV
  3. Nonlinearity. As a fraction of full scale: 457320=0.61%

Answer the full-scale output is 7320 µV, and the nonlinearity is 45 µV at 150 mmHg, 0.61% of full scale.

Worked example

An ECG front end has a gain of 1000 and feeds a 12-bit converter spanning ±5 V. What's its resolution at the electrodes? And what total error do independent errors of 1, 0.5 and 2 mmHg give?

Show the solution
  1. Step at the converter. The ±5 V span is 10 V, split into 4096 steps: q=10V4096=2.44mV
  2. Step at the electrodes. Dividing by the gain of 1000 gives qin=2.44μV fine for a 1 mV ECG.
  3. Combine the independent errors. They add as the root of the sum of squares: e=1+0.25+4=2.29mmHg dominated by the largest.

Answer the resolution is 2.44 µV at the electrodes (2.44 mV at the converter), and the three errors combine to 2.29 mmHg.

Worked example

A pressure transducer is calibrated at 0, 100, 200 and 300 mmHg and reads 0.02, 2.53, 5.01 and 7.55 mV. Find its sensitivity and offset by least squares, and the largest deviation from the fitted line.

Show the solution
  1. Means. x¯=150mmHgy¯=3.778mV
  2. Least-squares slope. S=∑(x−x¯)(y−y¯)∑(x−x¯)2=1253.550000=0.02507mV/mmHg
  3. Offset. b=y¯−Sx¯=0.017mV
  4. Residuals. The four points sit +0.003, +0.006, −0.021 and +0.012 mV off the line.
  5. Largest deviation. The worst point sits 0.021 mV, 0.8 mmHg, from the line: 0.0217.5=0.28% of the 7.5 mV span, the nonlinearity this calibration would quote.

Answer the sensitivity is 0.02507 mV/mmHg and the offset 0.017 mV, and the largest deviation is 0.021 mV (0.8 mmHg, 0.28% of the span).

Check

A pressure transducer's sensitivity is 10 µV per volt per mmHg, with 2.5 V excitation. What's its output at 250 mmHg?

Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.

Check

A sensor's output is Sx+ax2 with S = 20 and a = 0.001 (same units) over 0 to 300. What's its nonlinearity against the endpoint line, in percent of full scale?

Check

An amplifier with gain 5000 feeds a 10-bit converter spanning 5 V. What's one step, referred to the amplifier's input?

Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.

Check

A transducer's zero drifts by 0.2 mmHg per °C. How large an error does a 5 °C temperature change cause?

Check

A transducer's sensitivity falls by 0.002 (as a fraction) per °C. By what percent is a reading wrong after a 5 °C rise, whatever the pressure?

Check

A measurement has independent errors of 1, 0.3 and 3 mmHg. What's the combined error?

Check

A sensor with output impedance 350 Ω drives an amplifier whose input impedance is 1 kΩ.

(a)

By what percent does loading lower the measured signal?

(b)

What total resistance does the signal source see?

Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.

Check

A transducer's sensitivity S changes by a fraction k per °C and its zero by z output units per °C; the temperature moves by T from calibration while it measures x.

Step 1

What does it output?

Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.

Explain why

How do interfering and modifying inputs differ, and why does the difference matter for correcting them?

Practice

Keep practicing

10 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.