S15 Biomedical instrumentation · Practice
Medical measurement systems
Problems dealt one at a time, with new numbers every time. The kinds you miss come back more often, and you can keep going as long as you like.
Check
A pressure transducer's sensitivity is 10 µV per volt per mmHg, with 2.5 V excitation. What's its output at 250 mmHg?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
A sensor's output is with S = 20 and a = 0.001 (same units) over 0 to 300. What's its nonlinearity against the endpoint line, in percent of full scale?
Check
An amplifier with gain 5000 feeds a 10-bit converter spanning 5 V. What's one step, referred to the amplifier's input?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A transducer's zero drifts by 0.2 mmHg per °C. How large an error does a 5 °C temperature change cause?
Check
A transducer's sensitivity falls by 0.002 (as a fraction) per °C. By what percent is a reading wrong after a 5 °C rise, whatever the pressure?
Check
A measurement has independent errors of 1, 0.3 and 3 mmHg. What's the combined error?
Check
A sensor with output impedance 350 Ω drives an amplifier whose input impedance is 1 kΩ.
(a)
By what percent does loading lower the measured signal?
(b)
What total resistance does the signal source see?
Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.
Check
Output: 2 µV/V/mmHg, 1 V, 300 mmHg?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Input-referred step: 1 V, 8 bits, gain 10000?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
Combined error: 0.1, 0.1, 5?
A kind of problem counts as solid once you've got it right on the first try twice in a row. Back to the lesson · Practice every topic in S15