S15 Biomedical instrumentation · Practice
Midterm exam
Problems dealt one at a time, with new numbers every time. The kinds you miss come back more often, and you can keep going as long as you like.
Check
A 4 µV/V/mmHg transducer at 4 V; a gain of 300 feeding a 14-bit converter spanning 10 V.
(a)
Output at 200 mmHg?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Input-referred step?
(c)
That step in mmHg?
Check
A probe with τ = 6 s meets a 2 °C step; a catheter with = 50 Hz and ζ = 0.2.
(a)
Probe error after 30 s?
(b)
Catheter amplitude ratio at 10 Hz?
Check
Gauges with G = 2 at ±400 microstrain in a bridge at 10 V; separately, a quarter bridge with ΔR/R = 0.04.
(a)
Output with 4 active arms?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Exact quarter-bridge output?
Check
A capacitive sensor (0.5 cm², 0.05 mm, εr = 6); a piezoelectric sensor (d = 400 pC/N, 1 nF, read across 1 GΩ).
(a)
Capacitance?
Write units the way you normally would: 1.5 pF and 0.0015 nF are the same answer.
(b)
Initial piezo voltage for 0.5 N?
(c)
Piezo corner frequency?
Check
A thermistor of 8 kΩ at 25 °C with B = 3600 K, carrying 100 µA with dissipation constant 1 mW/°C.
(a)
Resistance at 40 °C?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
Self-heating there?
Check
Potassium 135 mM inside and 3.5 mM outside; sodium 12 mM inside and 145 mM outside; = 0.06; 37 °C.
(a)
Potassium's Nernst potential?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
GHK resting potential (K⁺ and Na⁺ only)?
Check
An axon 1 µm across with = 5000 Ω·cm² and = 150 Ω·cm; a conduction study over 30 cm with latencies 8 and 3 ms.
(a)
Length constant?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
(b)
Conduction velocity?
Check
A 2 × 10⁻⁵ A·m dipole in tissue of 0.35 S/m.
(a)
Potential 20 cm away at 25°?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
On the axis at the same distance?
Check
Lead I reads 0.35 mV and lead II 0.8 mV; R waves are 24 mm apart at 25 mm/s; QT = 0.4 s.
(a)
Electrical axis?
(b)
Heart rate?
(c)
Bazett's QTc?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
An electrode with = 0.5 kΩ, = 60 kΩ, = 0.3 µF; offsets up to 200 mV into a stage that swings ±4 V.
(a)
Impedance at 40 Hz?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
Largest first-stage gain?
Check
An instrumentation amplifier ( = 3 kΩ, = 25 kΩ, unity second stage, CMRR 90 dB, common-mode input impedance 100 MΩ) with 25 mV of common mode and 50 kΩ of electrode mismatch.
(a)
Gain?
(b)
Total input-referred common-mode error, adding both causes?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A 120 V, 50 Hz line couples through 4 pF to a patient on a driven-right-leg circuit ( = 8 MΩ, = 3 MΩ, = 25 kΩ).
(a)
Displacement current?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
Common-mode voltage?
Check
An evoked response of 0.5 µV in 10 µV of background; a 64-channel recorder at 500 per second with 16-bit samples.
(a)
Sweeps for an SNR of 8?
(b)
Data rate?
Write units the way you normally would: 1.5 Mb/s, 1500 kb/s and 0.0015 Gb/s are the same answer.
Check
A first-order high-pass at 0.05 Hz; a foil gauge (G = 2.1) with a 50 ppm/°C coefficient.
(a)
Phase lead at 0.5 Hz?
(b)
Apparent strain in microstrain for a 3 °C change?
Check
A pulse oximeter's 940 nm LED is switched at 1 kHz, and its photodiode (quantum efficiency 0.9) receives 2 µW of it. After amplification, the demodulator sees a 40 mV carrier at 20° to its reference.
(a)
Photocurrent from the LED's light?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
Average demodulator output?
Check
An isolated front end has a 5 pF converter and a 2 pF signal isolator across its barrier, with 120 dB of IMRR and a gain of 500; 230 V at 60 Hz stands across the barrier.
(a)
What leakage flows through the patient?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
How much line interference appears at the output?
A kind of problem counts as solid once you've got it right on the first try twice in a row. Back to the lesson · Practice every topic in S15