S15 · Lesson 43 · Pacemakers and defibrillators
Exciting tissue with current pulses and the strength–duration curve, chronaxie and the pulse width that costs least energy, pacing pulse energy and battery life, demand pacing, and defibrillators with their stored and delivered energy and discharge waveforms
Most of this course measures the body; this lesson acts on it. A pacemaker is a tiny pulse generator, sealed in titanium with a lithium battery, that must reliably excite heart muscle for a decade or more on about an ampere-hour of charge. A defibrillator does the opposite in scale: it dumps hundreds of joules in a few milliseconds to reset a heart in fibrillation. Both are designed around the same physiology from lesson 11: a cell fires when its membrane is depolarized past threshold, and a stimulus must deliver enough charge quickly enough to get it there.
- The strength–duration curve: the threshold current for a pulse of width t, , where is the rheobase (threshold for very long pulses) and c the chronaxie, the width at which twice the rheobase is needed.
- The energy per pulse into a fixed resistance goes as , which is smallest at t = c.
- Narrow pulses need large currents; wide ones waste charge while the membrane leaks.
Predict
A pacemaker delivers 2.5 V pulses 0.4 ms long into a 500 Ω lead. What energy does each pulse take?
= 6.25 × 0.0004/500 = 5 µJ, from a current of 5 mA and a charge of 2 µC. At 70 beats per minute that's under 6 µW of pacing, small next to the electronics' own drain, which is why a single battery can last well over a decade.
Name it
Strength and duration
A cathodic current pulse depolarizes the membrane by charging its capacitance while current also leaks away. The resulting threshold is well described by Weiss's law, linear in charge, , or in current
The energy into a resistance R, , is minimized at t = c, which is why pacemaker pulses are set near the chronaxie, typically a few tenths of a millisecond for heart muscle.
Pacemakers
A pacemaker delivers pulses of a few volts and a fraction of a millisecond through a lead to the heart, with the can or a second electrode as return. Modern demand pacemakers sense the heart's own activity and pace only when it's missing; a three-letter code names the chamber paced, the chamber sensed and the response (VVI paces and senses the ventricle and inhibits on a sensed beat; DDD works on both chambers). Pulse energy , charge , and battery life
where the average includes the pacing charge per second and the circuit's quiescent current.
Defibrillators
Ventricular fibrillation is chaotic electrical activity with no effective pumping. A defibrillator stores energy in a capacitor,
and discharges it through the chest (about 50 Ω between paddles) to depolarize most of the heart at once. External units store up to about 360 J at a few kilovolts; implanted ones (ICDs) deliver 30 to 40 J through internal leads. Modern waveforms are biphasic truncated exponentials: the capacitor discharges with τ = RC, is cut off after a few milliseconds, then reversed. A discharge truncated at t delivers the fraction of the stored energy.
Worked example
Estimate a pacemaker's battery life: 2 µC per pulse at 70 per minute, 5 µA of circuit current, a 1.2 Ah cell.
Show the solution
- Average pacing current.
- Total current. Adding the 5 µA of circuit current:
- Battery life. an upper bound, since real cells lose capacity and voltage near the end.
Answer the battery lasts at most about 18.7 years.
Worked example
A defibrillator charges 32 µF to 5 kV and discharges into 50 Ω. Find the stored energy, the time constant, the peak current, and the fraction delivered if the pulse is cut off after 4 ms.
Show the solution
- Stored energy.
- Time constant.
- Peak current.
- Fraction delivered by 4 ms. of the energy.
Answer it stores 400 J, with a 1.6 ms time constant and a 100 A peak, and cutting the pulse off at 4 ms still delivers 99.3% of the energy.
Worked example
An implantable defibrillator must store 30 J in a 150 µF capacitor. What voltage does that take, and how long does charging take from a battery through a converter drawing 10 W at 80% efficiency?
Show the solution
- Voltage.
- Charging time. At 80% efficiency the capacitor receives 8 W: during which the arrhythmia continues.
- What it means for the design. That delay, and the battery's ability to deliver 10 W at all, drive the design of the high-voltage section, from the battery chemistry to the transformer.
Answer the capacitor must be charged to 632 V, which takes 3.75 s.
Check
Tissue has a rheobase of 1 mA and a chronaxie of 0.3 ms. What threshold current does a 1 ms pulse need?
Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.
Check
A pacing pulse of 2.5 V lasts 0.2 ms into an 800 Ω lead. What energy does it deliver?
Write units the way you normally would: 1.5 µJ, 1500 nJ and 0.0015 mJ are the same answer.
Check
A pacemaker draws 8 µA continuously and delivers 2 µC per pulse at 60 pulses per minute from a 2.5 Ah battery. How many years does the battery last?
Check
A defibrillator charges a 32 µF capacitor to 0.7 kV. How much energy does it store?
Write units the way you normally would: 1.5 J, 1500 mJ and 0.0015 kJ are the same answer.
Check
A 32 µF capacitor discharges into 50 Ω and is cut off after 6 ms. What fraction of its stored energy is delivered?
Check
A defibrillator at 3 kV discharges into a 40 Ω chest. What's the peak current?
Write units the way you normally would: 1.5 A, 1500 mA and 0.0015 kA are the same answer.
Check
An ICD stores 35 J in a 100 µF capacitor, charged through a converter drawing 12 W at efficiency 0.85.
(a)
To what voltage must it charge?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
(b)
How long does charging take?
Check
A defibrillator charges C µF to V kV and discharges it into R ohms, truncating after t ms.
Step 1
How much energy does it store, in joules?
Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.
Step 2
What is the peak current, in amperes?
Step 3
What fraction of the energy has it delivered by truncation?
Explain why
Why is there a pulse width that minimizes the energy needed to stimulate tissue?
To fire, the membrane must be charged by a threshold amount, but some of the injected charge leaks across the membrane while the pulse is on. A very short pulse wastes little charge to leakage, but delivering the charge so quickly requires a large current, and the energy dissipated in the electrode and tissue grows as current squared. A very long pulse needs only the rheobase current, but it keeps driving current long after it could have finished, so the energy grows with duration. In between, the energy, proportional to , has a minimum exactly at the chronaxie, which is why it's the natural choice for battery-powered stimulators.