S15 Biomedical instrumentation · Practice
Pacemakers and defibrillators
Problems dealt one at a time, with new numbers every time. The kinds you miss come back more often, and you can keep going as long as you like.
Check
Tissue has a rheobase of 1 mA and a chronaxie of 0.3 ms. What threshold current does a 1 ms pulse need?
Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.
Check
A pacing pulse of 2.5 V lasts 0.2 ms into an 800 Ω lead. What energy does it deliver?
Write units the way you normally would: 1.5 µJ, 1500 nJ and 0.0015 mJ are the same answer.
Check
A pacemaker draws 8 µA continuously and delivers 2 µC per pulse at 60 pulses per minute from a 2.5 Ah battery. How many years does the battery last?
Check
A defibrillator charges a 32 µF capacitor to 0.7 kV. How much energy does it store?
Write units the way you normally would: 1.5 J, 1500 mJ and 0.0015 kJ are the same answer.
Check
A 32 µF capacitor discharges into 50 Ω and is cut off after 6 ms. What fraction of its stored energy is delivered?
Check
A defibrillator at 3 kV discharges into a 40 Ω chest. What's the peak current?
Write units the way you normally would: 1.5 A, 1500 mA and 0.0015 kA are the same answer.
Check
An ICD stores 35 J in a 100 µF capacitor, charged through a converter drawing 12 W at efficiency 0.85.
(a)
To what voltage must it charge?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
(b)
How long does charging take?
Check
Threshold: rheobase 0.1 mA, chronaxie 0.1 ms, width 5 ms?
Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.
Check
Stored energy: 10 µF at 0.5 kV?
Write units the way you normally would: 1.5 J, 1500 mJ and 0.0015 kJ are the same answer.
Check
Delivered fraction: 25 Ω, 20 µF, cut at 10 ms?
A kind of problem counts as solid once you've got it right on the first try twice in a row. Back to the lesson · Practice every topic in S15