A thermistor of 5 kΩ at 25 °C with B = 3500 K, at 39 °C.
(a)
Resistance?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
Coefficient in %/°C?
(c)
About how many ohms per °C?
Check
A platinum RTD with = 100 Ω; a thermocouple of 50 µV/°C.
(a)
RTD resistance at 30 °C?
Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.
(b)
Thermocouple voltage for a 25 °C difference?
Check
A 4 kΩ thermistor carrying 20 µA with dissipation constant 4 mW/°C; a body at 40 °C.
(a)
Self-heating?
(b)
Peak emission wavelength?
Write units the way you normally would: 1.5 µm, 1500 nm and 0.0015 mm are the same answer.
Check
Potassium 130 mM inside and 4 mM outside; calcium 2.5 mM outside and 0.0003 mM inside; 37 °C.
(a)
Potassium's Nernst potential?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Calcium's?
Check
A K⁺/Na⁺ membrane with = 0.02 (140/3.5 mM K⁺, 12/145 mM Na⁺, inside/outside); separately, conductances 1.5, 0.3, 0.2 for K⁺, Na⁺, Cl⁻ with Nernst potentials −95, +66.6 and −90 mV.
(a)
GHK potential?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Parallel-conductance potential?
Check
A membrane of 1 µF/cm² and area 10 × 10⁻⁵ cm² swings by 80 mV.
(a)
Charge per area, in µC/cm²?
Write units the way you normally would: 1.5 µC/cm², 1500 nC/cm² and 0.0015 mC/cm² are the same answer.
(b)
Number of monovalent ions?
Check
An axon 1 µm across, = 4000 Ω·cm², = 200 Ω·cm, = 1.2 µF/cm².
(a)
Length constant?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
(b)
Time constant?
(c)
What's left of a steady 15 mV at 0.5 mm?
Check
A conduction study with sites 20 cm apart and latencies 10 and 3 ms; a myelinated fiber 12 µm across.
(a)
Conduction velocity?
Write units the way you normally would: 1.5 m/s, 1500 mm/s and 0.0015 km/s are the same answer.
(b)
Travel time over 0.5 m for the fiber?
Check
A 1.5 × 10⁻⁵ A·m dipole in tissue of 0.3 S/m.
(a)
Potential 25 cm away at 20°?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
At twice the distance?
Check
Lead I reads 0.4 mV and lead II 0.6 mV.
(a)
Lead III?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
aVL?
(c)
Electrical axis?
Check
A heart vector of 0.7 mV at 10°.
(a)
Lead I?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
The augmented lead along 90°?
Check
R waves 15 mm apart at 25 mm/s, with a QT of 0.42 s.
(a)
Heart rate per minute?
(b)
Bazett's corrected QT?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
Extracellular potassium rises from 4 to 7 mM; K⁺ is 140 mM inside, Na⁺ 12/145 mM and Cl⁻ 4/120 mM (inside/outside), with = 1 : 0.05 : 0.45, at 37 °C.
(a)
How far does the potassium Nernst potential rise?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
How far does the resting potential rise?
Check
A Holter recorder stores 3 channels at 250 samples a second and 16 bits for 24 hours, on 32 MB of memory.
(a)
How much raw data does a day produce?
(b)
What compression ratio fits it?
Check
A 5 kΩ thermistor (at 25 °C, B = 3400 K) is to be linearized around 37 °C with a parallel resistor.
(a)
What resistor puts the inflection at 37 °C?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
A kind of problem counts as solid once you've got it right on the first try twice in a row. Back to the lesson · Practice every topic in S15