S15 · Problem set 5
Problem set 5: Goldman and Nernst potentials, membrane charge, cables and conduction, dipoles in a volume conductor, limb leads and the axis, electrode impedance and noise, instrumentation and isolation amplifiers, the driven right leg, floating monitors, averaging and spike detection
These problems cover lessons 10 to 19, from the ion gradients that make biopotentials to the amplifiers that record them: membranes and cables, the ECG's leads, electrodes and their noise, instrumentation and isolation amplifiers, interference and its rejection, and the signal processing that pulls small signals out of noise.
Check
A membrane has = 1 : 0.05 : 0.6, with K⁺ 140/4 mM, Na⁺ 12/135 mM and Cl⁻ 4/120 mM (inside/outside), at 37 °C.
(a)
What is its resting potential?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
How far is it below the sodium Nernst potential (the driving force on Na⁺)?
Check
Calcium is 1.2 mM outside a cell and 0.0001 mM inside; the membrane sits at −60 mV, at 37 °C.
(a)
What is calcium's Nernst potential?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
What driving force pushes calcium inward?
Check
An action potential swings a membrane of 1 µF/cm² by 100 mV, in a cell 10 µm across (a sphere).
(a)
How many monovalent ions per square centimeter must cross?
(b)
How many ions in all, for the whole cell?
Check
An unmyelinated fiber 2 µm across has = 5000 Ω·cm² and = 150 Ω·cm; it's 4 mm long and sealed at its far end.
(a)
What is its length constant?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
(b)
What fraction of a steady voltage held at one end reaches the sealed end?
Check
A nerve's myelinated fibers range from 2 to 9 µm across, at 6 m/s per micrometer; a recording site is 0.6 m from the stimulus.
(a)
When does the response begin?
Write units the way you normally would: 1.5 ms, 1500 µs and 0.0015 s are the same answer.
(b)
Over how long is the volley dispersed?
Check
A current dipole of 1 × 10⁻⁵ A·m sits in tissue of 0.2 S/m.
(a)
What potential does it produce 20 cm away at 70° to its axis?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Lead I reads 0.4 mV and lead II 0.3 mV; R waves are 30 mm apart at 25 mm/s, and the QT interval is 0.4 s.
(a)
What is the electrical axis?
(b)
What is Bazett's corrected QT?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
An electrode has = 0.5 kΩ, = 20 kΩ and = 2 µF.
(a)
What is its impedance's magnitude at 60 Hz?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
Check
An instrumentation amplifier has = 5 kΩ, = 25 kΩ and a second stage of gain 50, with 90 dB of CMRR; 10 mV of common mode is present.
(a)
What is its gain?
(b)
What input-referred error does the common mode cause?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
An amplifier's noise is 5 nV/√Hz and 0.05 pA/√Hz, and each input sees 200 kΩ at 37 °C, over 500 Hz.
(a)
What total input noise results?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
An isolated input's barrier has 20 pF and 140 dB of IMRR; 120 V at 60 Hz stands across it.
(a)
What leakage flows?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
What input-referred error does the isolation-mode voltage cause?
Check
A driven-right-leg circuit has = 1 MΩ, = 2 MΩ and = 50 kΩ, with 0.5 µA of displacement current.
(a)
What effective resistance ties the patient to the circuit?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
What common-mode voltage results?
Check
A battery-powered monitor's common has 100 pF to earth; the patient couples 1 pF to a 230 V, 60 Hz line, and the reference electrode is 100 kΩ.
(a)
What common-mode voltage reaches the amplifier?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
An evoked response of 1 µV sits in 10 µV RMS of EEG, recorded by a 24-bit converter spanning ±2 V after a gain of 1000.
(a)
How many sweeps reach a signal-to-noise ratio of 10?
(b)
What step does the converter resolve at the electrodes?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A spike detector's noise is flat from 0 to 3000 Hz, with its threshold at 3.5 standard deviations; the neuron fires 50 times a second.
(a)
How often does noise alone cross the threshold?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
What fraction of detected events are false?
Have paper and a calculator ready. Work each problem out by hand, then enter the answer. Your first try on each one counts toward your score.
Practice
Keep practicing
15 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.