S15 Biomedical instrumentation · Practice
Problem set 5: biopotentials, electrodes and amplifiers
Problems dealt one at a time, with new numbers every time. The kinds you miss come back more often, and you can keep going as long as you like.
Check
A membrane has = 1 : 0.05 : 0.6, with K⁺ 140/4 mM, Na⁺ 12/135 mM and Cl⁻ 4/120 mM (inside/outside), at 37 °C.
(a)
What is its resting potential?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
How far is it below the sodium Nernst potential (the driving force on Na⁺)?
Check
Calcium is 1.2 mM outside a cell and 0.0001 mM inside; the membrane sits at −60 mV, at 37 °C.
(a)
What is calcium's Nernst potential?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
What driving force pushes calcium inward?
Check
An action potential swings a membrane of 1 µF/cm² by 100 mV, in a cell 10 µm across (a sphere).
(a)
How many monovalent ions per square centimeter must cross?
(b)
How many ions in all, for the whole cell?
Check
An unmyelinated fiber 2 µm across has = 5000 Ω·cm² and = 150 Ω·cm; it's 4 mm long and sealed at its far end.
(a)
What is its length constant?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
(b)
What fraction of a steady voltage held at one end reaches the sealed end?
Check
A nerve's myelinated fibers range from 2 to 9 µm across, at 6 m/s per micrometer; a recording site is 0.6 m from the stimulus.
(a)
When does the response begin?
Write units the way you normally would: 1.5 ms, 1500 µs and 0.0015 s are the same answer.
(b)
Over how long is the volley dispersed?
Check
A current dipole of 1 × 10⁻⁵ A·m sits in tissue of 0.2 S/m.
(a)
What potential does it produce 20 cm away at 70° to its axis?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Lead I reads 0.4 mV and lead II 0.3 mV; R waves are 30 mm apart at 25 mm/s, and the QT interval is 0.4 s.
(a)
What is the electrical axis?
(b)
What is Bazett's corrected QT?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
An electrode has = 0.5 kΩ, = 20 kΩ and = 2 µF.
(a)
What is its impedance's magnitude at 60 Hz?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
Check
An instrumentation amplifier has = 5 kΩ, = 25 kΩ and a second stage of gain 50, with 90 dB of CMRR; 10 mV of common mode is present.
(a)
What is its gain?
(b)
What input-referred error does the common mode cause?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
An amplifier's noise is 5 nV/√Hz and 0.05 pA/√Hz, and each input sees 200 kΩ at 37 °C, over 500 Hz.
(a)
What total input noise results?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
An isolated input's barrier has 20 pF and 140 dB of IMRR; 120 V at 60 Hz stands across it.
(a)
What leakage flows?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
What input-referred error does the isolation-mode voltage cause?
Check
A driven-right-leg circuit has = 1 MΩ, = 2 MΩ and = 50 kΩ, with 0.5 µA of displacement current.
(a)
What effective resistance ties the patient to the circuit?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
What common-mode voltage results?
Check
A battery-powered monitor's common has 100 pF to earth; the patient couples 1 pF to a 230 V, 60 Hz line, and the reference electrode is 100 kΩ.
(a)
What common-mode voltage reaches the amplifier?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
An evoked response of 1 µV sits in 10 µV RMS of EEG, recorded by a 24-bit converter spanning ±2 V after a gain of 1000.
(a)
How many sweeps reach a signal-to-noise ratio of 10?
(b)
What step does the converter resolve at the electrodes?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A spike detector's noise is flat from 0 to 3000 Hz, with its threshold at 3.5 standard deviations; the neuron fires 50 times a second.
(a)
How often does noise alone cross the threshold?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
What fraction of detected events are false?
A kind of problem counts as solid once you've got it right on the first try twice in a row. Back to the lesson · Practice every topic in S15