S15 · Lesson 33 · Pulse oximetry
Beer–Lambert absorption, the absorption spectra of oxygenated and deoxygenated hemoglobin, the photoplethysmogram's pulsatile and steady parts, the ratio of ratios at red and infrared, why oximeters are calibrated empirically, and oxygen content
A clip on the finger reads the oxygen saturation of arterial blood, continuously and without a needle, from two LEDs and a photodiode. Hemoglobin changes color when it picks up oxygen: deoxygenated blood absorbs red light strongly and oxygenated blood absorbs more in the near infrared. The difficulty is that the finger is mostly skin, bone and venous blood, which also absorb. The pulse oximeter's insight is to use only the part of the absorption that pulses with the heartbeat, because only arterial blood pulses.
- The ratio of ratios R, the red pulsatile-to-steady ratio over the infrared one, against saturation. The solid curve is Beer–Lambert theory with hemoglobin's extinction coefficients at 660 and 940 nm; the dashed line is the common empirical calibration SpO₂ ≈ 110 − 25R.
- They disagree badly: light scattered in tissue travels different path lengths at the two wavelengths, which the simple theory ignores.
- That's why every oximeter is calibrated against arterial blood samples from volunteers.
Predict
An oximeter measures red AC/DC = 0.02/1.0 and infrared AC/DC = 0.03/1.2. Using SpO₂ = 110 − 25R, what does it display?
R = (0.02/1.0)/(0.03/1.2) = 0.02/0.025 = 0.8, so SpO₂ = 110 − 25 × 0.8 = 90%. Dividing each wavelength's pulsatile signal by its own steady level cancels the LED brightness, detector gain and the tissue's steady absorption, leaving a ratio that depends on the arterial blood's color.
Name it
Beer–Lambert absorption
Light of intensity through a path l of an absorber at concentration c leaves
with ε the molar extinction coefficient. The absorbance adds over absorbers. Approximate extinction coefficients (in cm⁻¹/M, from Prahl's tabulation): at 660 nm, 3227 for deoxyhemoglobin and 320 for oxyhemoglobin; at 940 nm, 693 and 1214.
The photoplethysmogram
Each heartbeat swells the arterioles slightly, lengthening the blood path by Δl and dimming the transmitted light. The detected signal has a steady part (DC) from tissue, venous blood and the non-pulsing arterial blood, and a small pulsatile part (AC), typically 0.5 to 5% of the DC. For small changes, .
The ratio of ratios
In the Beer–Lambert model, with saturation S,
Real tissue scatters light heavily and the effective path lengths differ at the two wavelengths, so devices use an empirical calibration, often close to SpO₂ ≈ 110 − 25R over the clinical range. Accuracy is about ±2 to 3% above 70% saturation; motion, poor perfusion, nail polish, ambient light, and carbon monoxide (carboxyhemoglobin looks like oxyhemoglobin in red) degrade it.
Oxygen content
Saturation isn't the whole story. Each gram of hemoglobin carries about 1.34 mL of O₂ when saturated, and plasma dissolves 0.003 mL/dL per mmHg:
Worked example
What does the Beer–Lambert model predict for R at 97% and 85% saturation, and what does the empirical line give?
Show the solution
- Theory at 97%.
- Theory at 85%. The same formula gives R = 0.666.
- Empirical line. At 97% it gives and at 85% it gives 1.0.
- Why they differ. The theory is off by tens of percent because it ignores scattering, which is the reason for empirical calibration.
Answer the theory predicts R = 0.340 at 97% and 0.666 at 85%; the empirical line gives 0.52 and 1.0.
Worked example
Blood has 15 g/dL of hemoglobin, 98% saturated, with a of 100 mmHg. What's its oxygen content? And what fraction of light does an absorbance of 0.4 transmit?
Show the solution
- Oxygen content. almost all carried by hemoglobin.
- Transmitted light.
Answer the blood carries 20.0 mL/dL of oxygen, and an absorbance of 0.4 transmits 0.398 of the light.
Worked example
A poorly perfused finger gives an infrared perfusion index (AC/DC) of 0.2%, with the DC level at 80% of a 12-bit converter's range. How many counts does the pulse span, and what would a 16-bit converter give?
Show the solution
- DC level in counts.
- Pulse span. so the waveform would be a staircase a few steps tall, and the ratio of ratios would carry about 15% quantization error.
- With 16 bits. A 16-bit converter gives 105 counts.
- What oximeters do instead. They subtract the DC level in hardware and amplify what's left before converting, which is what lets them read through weak perfusion.
Answer the pulse spans 6.6 counts on the 12-bit converter and 105 counts on a 16-bit one.
Check
Red AC/DC is 0.02/0.8 and infrared AC/DC is 0.05/1.5. What's the ratio of ratios?
Check
An oximeter calibrated as SpO₂ = 110 − 25R measures R = 0.7. What does it display, in percent?
Check
Using extinction coefficients of 3227 and 320 (Hb, HbO₂) at 660 nm and 693 and 1214 at 940 nm, what R does the Beer–Lambert model predict at 85% saturation?
Check
Light passes 0.5 cm through hemoglobin at 0.002 M with an extinction coefficient of 1000 cm⁻¹/M. What fraction is transmitted?
Check
Blood has 12 g/dL of hemoglobin at saturation 0.8 and a of 100 mmHg. What's its oxygen content?
Write units the way you normally would: 1.5 mL/dL, 1500 µL/dL and 0.0015 L/dL are the same answer.
Check
A sample transmits a fraction 0.2 of the incident light. What's its absorbance?
Check
An oximeter's infrared perfusion index is 0.5%, with the DC level at 0.5 of a 16-bit converter's range.
(a)
How many counts does the pulse span?
Check
An oximeter measures red AC a and DC b, and infrared AC c and DC d.
Step 1
What is the ratio of ratios?
Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.
Step 2
What saturation does the empirical calibration 110 − 25R give, in percent?
Explain why
Why does a pulse oximeter use only the pulsatile part of the light signal?
Light through a finger is absorbed by skin, bone, tissue and venous blood as well as arterial blood, and those steady absorbers vary from person to person and dominate the signal. Only the arterial blood volume changes with each heartbeat, so the pulsatile part of the transmitted light comes from arterial blood alone. Dividing that AC part by the DC level at the same wavelength removes the effects of LED brightness, detector sensitivity and the steady tissue absorption, and taking the ratio of the two wavelengths' normalized pulsations cancels the unknown size of the pulse. What's left depends mainly on the arterial blood's relative absorption at red and infrared, which is set by its oxygen saturation.