S15 Biomedical instrumentation · Practice
Pulse oximetry
Problems dealt one at a time, with new numbers every time. The kinds you miss come back more often, and you can keep going as long as you like.
0 solved
0 in a row
0 of 10 kinds solid
Check
Red AC/DC is 0.02/0.8 and infrared AC/DC is 0.05/1.5. What's the ratio of ratios?
Check
An oximeter calibrated as SpO₂ = 110 − 25R measures R = 0.7. What does it display, in percent?
Check
Using extinction coefficients of 3227 and 320 (Hb, HbO₂) at 660 nm and 693 and 1214 at 940 nm, what R does the Beer–Lambert model predict at 85% saturation?
Check
Light passes 0.5 cm through hemoglobin at 0.002 M with an extinction coefficient of 1000 cm⁻¹/M. What fraction is transmitted?
Check
Blood has 12 g/dL of hemoglobin at saturation 0.8 and a of 100 mmHg. What's its oxygen content?
Write units the way you normally would: 1.5 mL/dL, 1500 µL/dL and 0.0015 L/dL are the same answer.
Check
A sample transmits a fraction 0.2 of the incident light. What's its absorbance?
Check
An oximeter's infrared perfusion index is 0.5%, with the DC level at 0.5 of a 16-bit converter's range.
(a)
How many counts does the pulse span?
Check
R: red 0.005/0.5, infrared 0.08/2?
Check
Beer–Lambert R at 80%?
Check
Oxygen content: Hb 6 g/dL, S = 0.6, = 500 mmHg?
Write units the way you normally would: 1.5 mL/dL, 1500 µL/dL and 0.0015 L/dL are the same answer.
A kind of problem counts as solid once you've got it right on the first try twice in a row. Back to the lesson · Practice every topic in S15