S15 · Lesson 45 · Therapeutic devices: electrosurgery, dialysis, ventilators and infusion pumps
Cutting and coagulating with radio-frequency current and keeping the return electrode cool, the artificial kidney's clearance and Kt/V, a ventilator's airway pressure from resistance and compliance, stepper-driven infusion pumps, and a short tour of lithotripters, incubators, lasers and assist devices
Most of this course has been about measuring the body. The devices in this lesson act on it. A surgeon cuts with a current of half an ampere that the patient can't feel. A dialysis machine does a kidney's filtering three times a week. A ventilator breathes for a patient who can't, and an infusion pump delivers a drug a fraction of a milliliter at a time for days. Each is an engineered system with a physical model at its core, and each can hurt the patient when the model's assumptions fail.
- A ventilator delivers a volume-controlled breath: a constant flow Q = 0.5 L/s for 1 s (a 500 mL tidal volume), then lets the patient exhale passively against the set end-expiratory pressure (PEEP).
- The lungs are modeled as an airway resistance R in series with a compliance C. The solid curve is the pressure the ventilator sees at the airway; the dashed one is the pressure in the alveoli.
- At the start of the breath the airway pressure jumps by RQ, the drop across the airways, then climbs as the lungs fill. Stiff lungs (small C) raise the peak; narrowed airways (large R) raise the gap between airway and alveolar pressure and slow the exhalation, whose time constant is RC.
Predict
An electrosurgical unit passes about 0.6 A of radio-frequency current through the patient to cut tissue. Why doesn't that current stop the heart, when 0.1 A at 60 Hz can?
Excitable membranes integrate current: lesson 43's strength–duration curve shows that short pulses need much more current to stimulate. A half-cycle at 500 kHz lasts 1 µs, so the threshold rises far above any current used, and the only effect left is heating. Electrosurgery runs from about 250 kHz to a few megahertz for exactly this reason.
Name it
Electrosurgery
An electrosurgical unit drives radio-frequency current from a small active electrode, where the current density is high enough to heat tissue rapidly, into a large dispersive (return) electrode on the patient's skin, where it spreads out and heats little. Webster gives the ranges: cutting uses continuous waves at 500 kHz to 2.5 MHz, open-circuit voltages up to 9 kV and 100 to 750 W, vaporizing cells in a spark; coagulation uses bursts of damped sinusoids repeated about 120 times a second, at 300 to 2000 V and 80 to 200 W into 500 Ω, drying tissue and sealing vessels; a blend combines them. Bipolar forceps keep the current between the two tips.
Heating per unit area at the return electrode goes as the current density squared. If a pad partly peels off, the current crowds into what's left, and halving the contact area quadruples the heating there. Return-electrode monitors split the pad in two and check the impedance between the halves continuously. Stray capacitance from cables and the patient to ground gives the current other paths at radio frequencies, which is why burns can appear at ECG electrodes or wherever skin touches a grounded surface.
Hemodialysis
An artificial kidney passes blood on one side of a semipermeable membrane, today usually 10,000 to 15,000 hollow fibers about 0.2 mm across, and a dialysate on the other. Small waste molecules such as urea diffuse across. A dialyzer's performance is its clearance, the blood flow it cleans completely:
Treating urea as spread through a volume V of body water (about 58% of body weight), its concentration falls as over a session. The dose is stated as Kt/V, with about 1.2 per session the usual minimum for three sessions a week, and the urea reduction ratio is . The machine monitors the dialysate's conductivity (its electrolyte concentration) and temperature, blood-leak and air detectors, and the pressures that drive ultrafiltration, the removal of excess water.
Ventilators
A ventilator pushes gas into the lungs, which behave to first order like a resistance R (the airways) in series with a compliance C (the lungs and chest wall). The airway pressure during a breath is
where Q is the flow and V the volume delivered so far. In volume control the ventilator sets the flow and tidal volume and the pressure follows; in pressure control it sets the pressure and the volume follows from the lungs' mechanics. PEEP, a few cmH₂O kept at the end of exhalation, holds small airways open. Exhalation is passive and decays with time constant RC, so about 3RC is needed to empty; too short an exhalation traps gas. Tidal volumes are now set by body size, about 6 to 8 mL per kilogram of ideal body weight, to avoid overstretching injured lungs. High-frequency ventilators move small volumes hundreds of times a minute.
Infusion pumps
Peristaltic pumps squeeze a tube with rollers; syringe pumps push a plunger with a lead screw. Both are driven by stepper motors, so each step delivers a fixed volume and the rate is set by the step frequency. They detect occlusion (rising line pressure) and air in line (by ultrasound) and stop with an alarm. Closed-loop systems adjust the rate to keep a measured variable, such as blood pressure or glucose, at a target.
Other therapeutic devices
- Lithotripters break kidney stones with focused shock waves: a source at one focus of an ellipsoidal reflector concentrates the pulse at the other focus, positioned on the stone. Water couples the shock into the body because soft tissue's acoustic impedance is nearly the same as water's (S16 covers why that matters).
- Infant incubators servo-control air or skin temperature, usually around 36 to 37 °C at the skin, and humidity.
- Therapeutic lasers cut, coagulate or ablate by choosing a wavelength the target absorbs strongly, such as CO₂ at 10.6 µm in water.
- Mechanical heart valves and ventricular assist devices replace or support the heart's pumping; their engineering problems are flow, wear and clotting.
Worked example
A volume-controlled breath delivers 500 mL at 0.5 L/s with PEEP = 5 cmH₂O to lungs with R = 10 cmH₂O·s/L and C = 50 mL/cmH₂O. Find the peak and plateau pressures and how long exhalation takes.
Show the solution
- Pressures the breath must supply. Pushing 0.5 L/s through the airways and filling the compliance with 0.5 L cost
- Peak pressure. PEEP plus both:
- Plateau pressure. During an inspiratory pause the flow stops and the resistive drop disappears: the alveolar pressure.
- What the two measure. The difference between peak and plateau measures resistance; the plateau minus PEEP measures compliance.
- Exhalation. so about 1.5 s is needed to exhale 95%.
- Time available. At 20 breaths a minute with 1 s of inspiration, the patient has 2 s: enough here, but not for a patient with asthma whose resistance is three times higher.
Answer the peak pressure is 20 cmH₂O and the plateau 15 cmH₂O, and exhaling 95% takes about 1.5 s.
Worked example
A dialyzer with 300 mL/min of blood flow lowers urea from 100 to 30 mg/dL across it. What dose does a 4-hour session give a 70 kg patient?
Show the solution
- Clearance.
- Volume cleared in the session. In 240 minutes:
- Kt/V. Urea spreads through about 0.58 × 70 = 40.6 L of body water: just above the usual minimum of 1.2.
- Urea reduction. The urea concentration at the end is = 29% of the start, a urea reduction ratio of 71%.
- Limits of the estimate. This single-pool estimate ignores urea made during the session and the rebound from tissues afterward, which clinical formulas correct for.
Answer the session gives Kt/V = 1.24, just above the usual minimum of 1.2, with a urea reduction ratio of 71%.
Worked example
An electrosurgical unit returns 1 A RMS through a 100 cm² dispersive pad. What current density does the skin see, and how much faster does it heat if the pad peels back to 20 cm² of contact?
Show the solution
- Current density on the full pad. low enough that blood flow carries the heat away.
- On 20 cm².
- Heating rate. Heating goes as the square of current density: so the skin under the remaining contact heats 25 times faster, enough to burn.
- Protection. That's why modern units monitor the pad's contact impedance with split electrodes and stop if it rises.
Answer the skin sees 10 mA/cm² on the full pad; peeled back to 20 cm², it sees 50 mA/cm² and heats 25 times faster.
Check
A syringe pump pushes a 1.5 cm diameter plunger with a lead screw of 0.5 mm pitch, turned by a stepper of 400 steps per revolution. At what step rate does it deliver 20 mL/h?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
Blood enters a dialyzer at 300 mL/min with urea at 80 mg/dL and leaves at 45 mg/dL. What's the dialyzer's clearance, in mL/min?
Write units the way you normally would: 1.5 mL/min, 1500 µL/min and 0.0015 L/min are the same answer.
Check
A dialyzer clears 220 mL/min for 3 hours on a 90 kg patient (body water 58% of weight). What's Kt/V?
Check
A session delivers Kt/V = 1.2. Ignoring urea generation, what's the urea reduction ratio?
Check
A ventilator delivers 0.5 L at 0.5 L/s with PEEP 5 cmH₂O to lungs with R = 20 cmH₂O·s/L and C = 0.07 L/cmH₂O. What's the peak airway pressure, in cmH₂O?
Check
Lungs with R = 15 cmH₂O·s/L and C = 0.03 L/cmH₂O exhale passively. About how long does it take to exhale 95% of the tidal volume?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
An electrosurgical unit sends 200 W through a 50 cm² return electrode. If only a fraction 0.5 of the pad stays in contact, by what factor does the heating per square centimeter under it rise?
Check
An electrosurgical unit returns 1 A RMS through a pad of 60 cm²; partial peeling leaves 30 cm² in contact.
(a)
What current density does the full pad carry?
Write units the way you normally would: 1.5 mA/cm^2, 1500 µA/cm^2 and 0.0015 A/cm^2 are the same answer.
(b)
By what factor does the heating rate rise after peeling?
Check
A syringe pump drives a plunger of diameter d (cm) with a lead screw of pitch p (mm), turned by a stepper with s steps per revolution at r steps per second.
Step 1
What volume does one step deliver, in milliliters?
Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.
Step 2
What flow results, in mL per hour?
Explain why
Why do electrosurgical units run at hundreds of kilohertz, and why does that create burn risks that a 60 Hz device wouldn't?
Nerve and muscle membranes respond to the charge delivered in a pulse, and a half-cycle at 500 kHz lasts only a microsecond, so the stimulation threshold at that frequency is far above the currents used. The current can therefore heat and cut without causing shocks, muscle contractions or fibrillation. The cost of the high frequency is that small capacitances become low impedances: a 100 pF path from the patient to a grounded table or monitor lead has about 3 kΩ of reactance at 500 kHz, against 27 MΩ at 60 Hz. Current can return through those paths instead of the dispersive pad, and wherever it leaves the body through a small area, such as an ECG electrode or a fingertip touching metal, the current density is high and burns result. Isolated generator outputs, return-electrode monitoring and careful placement of the pad reduce these risks.