S15 Biomedical instrumentation · Practice
Therapeutic devices: electrosurgery, dialysis, ventilators and infusion pumps
Problems dealt one at a time, with new numbers every time. The kinds you miss come back more often, and you can keep going as long as you like.
Check
A syringe pump pushes a 1.5 cm diameter plunger with a lead screw of 0.5 mm pitch, turned by a stepper of 400 steps per revolution. At what step rate does it deliver 20 mL/h?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
Blood enters a dialyzer at 300 mL/min with urea at 80 mg/dL and leaves at 45 mg/dL. What's the dialyzer's clearance, in mL/min?
Write units the way you normally would: 1.5 mL/min, 1500 µL/min and 0.0015 L/min are the same answer.
Check
A dialyzer clears 220 mL/min for 3 hours on a 90 kg patient (body water 58% of weight). What's Kt/V?
Check
A session delivers Kt/V = 1.2. Ignoring urea generation, what's the urea reduction ratio?
Check
A ventilator delivers 0.5 L at 0.5 L/s with PEEP 5 cmH₂O to lungs with R = 20 cmH₂O·s/L and C = 0.07 L/cmH₂O. What's the peak airway pressure, in cmH₂O?
Check
Lungs with R = 15 cmH₂O·s/L and C = 0.03 L/cmH₂O exhale passively. About how long does it take to exhale 95% of the tidal volume?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
An electrosurgical unit sends 200 W through a 50 cm² return electrode. If only a fraction 0.5 of the pad stays in contact, by what factor does the heating per square centimeter under it rise?
Check
An electrosurgical unit returns 1 A RMS through a pad of 60 cm²; partial peeling leaves 30 cm² in contact.
(a)
What current density does the full pad carry?
Write units the way you normally would: 1.5 mA/cm^2, 1500 µA/cm^2 and 0.0015 A/cm^2 are the same answer.
(b)
By what factor does the heating rate rise after peeling?
Check
Dialyzer clearance in mL/min: 260 mL/min, urea 66 in, 30 out?
Write units the way you normally would: 1.5 mL/min, 1500 µL/min and 0.0015 L/min are the same answer.
Check
Peak pressure, cmH₂O: 0.48 L at 0.37 L/s, R = 22, C = 0.093, PEEP 1.5?
Check
Urea reduction ratio for Kt/V = 1.2?
A kind of problem counts as solid once you've got it right on the first try twice in a row. Back to the lesson · Practice every topic in S15