S15 · Lesson 40 · Ultrasound imaging
Acoustic impedance and the echoes at tissue boundaries, ranging by echo time, frequency-dependent attenuation and the depth it allows, axial and lateral resolution, pulse rate and frame rate limits, and the quarter-wave matching layer on the transducer
An ultrasound scanner is a radar for soft tissue, built from the piezoelectric transducers of lesson 4. It sends a short burst of a few megahertz into the body, listens for echoes, and turns their delays into depths and their strengths into brightness. It's cheap, portable and uses no ionizing radiation, which is why it's the first imaging test for a pregnancy or a painful gallbladder. Its trade-offs are the ones every echo system faces: higher frequencies resolve finer detail but are absorbed faster, and every line of the image costs a round-trip time.
- The round-trip attenuation of an echo against depth, in decibels, with α ≈ 0.5 dB/(cm·MHz) for soft tissue.
- A scanner with 60 dB of usable dynamic range can image until the round trip costs 60 dB: about 12 cm at 5 MHz, 30 cm at 2 MHz.
- Higher frequency buys resolution: the readouts show the wavelength and the axial resolution for a two-cycle pulse.
Predict
Soft tissue has an acoustic impedance of 1.63 MRayl and air about 0.0004 MRayl. What fraction of ultrasound intensity reflects at a tissue–air boundary?
= 0.999. Essentially nothing crosses into air, which is why gel couples the probe to the skin and why ultrasound can't see past lung or bowel gas. Fat against soft tissue reflects only 0.7%, enough to see without blocking the beam.
Name it
Impedance and echoes
A medium's acoustic impedance is ; soft tissue is about 1060 kg/m³ × 1540 m/s = 1.63 MRayl, fat 1.38, bone about 7.8, air 0.0004. At a boundary hit squarely, the reflected fraction of intensity is
and goes on. It's the transmission-line reflection coefficient of C10, squared for power.
Ranging and timing
Scanners assume c = 1540 m/s everywhere, so an echo after time t comes from depth . The next pulse must wait for echoes from the deepest point, so the pulse repetition frequency is at most , and an image of N lines takes : the frame rate is .
Attenuation
Soft tissue absorbs and scatters about 0.5 dB per cm per MHz, roughly proportional to frequency, so the round trip to depth d costs dB. Time-gain compensation raises the receiver gain with time to even out the image.
Resolution
- Axial (along the beam): half the spatial pulse length, for an n-cycle pulse, with λ = c/f.
- Lateral: the beam width, about λ times the focal ratio (focal length over aperture) at the focus.
Both improve with frequency, which is limited by the depth needed: 2 to 5 MHz for abdomen and heart, 7 to 15 MHz for shallow structures.
Matching layers
PZT ceramic has an impedance of about 30 MRayl, so most energy would reflect at its face. A quarter-wavelength layer with matches them, just as a quarter-wave transformer matches transmission lines.
Worked example
An echo returns after 130 µs. How deep was the reflector? How much is a 5 MHz echo from 10 cm attenuated, and what's the axial resolution of a two-cycle 5 MHz pulse?
Show the solution
- Depth. The echo travels there and back:
- Attenuation. The round trip costs
- Wavelength.
- Axial resolution.
Answer the reflector was 10.0 cm deep, the echo is attenuated by 50 dB, and the axial resolution is 0.31 mm.
Worked example
What frame rate does a 128-line image to 15 cm allow? And what impedance and thickness should a 5 MHz matching layer have, if sound travels at 3000 m/s in it?
Show the solution
- Time per line.
- Frame rate. 128 lines take 25 ms, so the frame rate is 40 frames per second.
- Matching-layer impedance.
- Matching-layer thickness. A quarter wavelength in the layer:
Answer the image allows 40 frames per second, and the matching layer should be 7.0 MRayl and 0.15 mm thick.
Worked example
An unfocused 5 MHz transducer is 10 mm across. How long is its near field, how fast does its beam spread beyond it, and what lateral resolution does focusing at f-number 5 give?
Show the solution
- Wavelength. In soft tissue:
- Near field. The beam stays about as wide as the transducer for
- Spread beyond it. The beam spreads at a half-angle of
- Focused lateral resolution. Focusing narrows the beam at the focus to about λ times the f-number: but only over a limited depth, which is why scanners focus electronically at several depths.
Answer the near field is 81 mm long, the beam spreads at 2.2° beyond it, and focusing at f-number 5 gives about 1.5 mm of lateral resolution.
Check
Ultrasound passes from a medium of 1.63 MRayl into one of 1.99 MRayl. What fraction of the intensity reflects?
Check
An echo arrives 65 µs after the pulse. How deep is the reflector?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
Check
Tissue attenuates by 0.5 dB/(cm·MHz). How much is a 2 MHz echo from 15 cm deep attenuated?
Check
What's the axial resolution of a 2-cycle pulse at 2 MHz in soft tissue?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
Check
An image of 64 lines reaches 15 cm deep. What's the highest possible frame rate, in frames per second?
Check
What impedance, in MRayl, should a quarter-wave layer have to match a 30 MRayl transducer to 1.5 MRayl tissue?
Check
An ultrasound transducer 6 mm across works at 5 MHz in soft tissue (1540 m/s).
(a)
How long is its near field?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
(b)
What lateral resolution does focusing at f-number 4 give?
Check
Sound travels at 1540 m/s in tissue.
Step 1
How deep, in centimeters, is a reflector whose echo returns after t µs?
Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.
Step 2
What is the highest pulse rate that lets echoes from d cm return before the next pulse?
Explain why
Why must an ultrasound system trade imaging depth against resolution?
Resolution along the beam is set by the pulse length, and a pulse of a given number of cycles is shorter at higher frequency, so higher frequencies resolve finer detail; the beam can also be focused more tightly. But tissue absorbs ultrasound at a rate roughly proportional to frequency, and an echo must survive the round trip with enough strength to rise above the receiver's noise. At a fixed dynamic range, the depth that can be imaged falls in inverse proportion to frequency. So deep organs are scanned at a few megahertz with coarser resolution, and superficial structures such as tendons, thyroid or blood vessels near the skin are scanned at 10 MHz or more with sub-millimeter detail.
Practice
Keep practicing
10 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.