S15 · Lesson 40 · Ultrasound imaging

Acoustic impedance and the echoes at tissue boundaries, ranging by echo time, frequency-dependent attenuation and the depth it allows, axial and lateral resolution, pulse rate and frame rate limits, and the quarter-wave matching layer on the transducer

An ultrasound scanner is a radar for soft tissue, built from the piezoelectric transducers of lesson 4. It sends a short burst of a few megahertz into the body, listens for echoes, and turns their delays into depths and their strengths into brightness. It's cheap, portable and uses no ionizing radiation, which is why it's the first imaging test for a pregnancy or a painful gallbladder. Its trade-offs are the ones every echo system faces: higher frequencies resolve finer detail but are absorbed faster, and every line of the image costs a round-trip time.

  • The round-trip attenuation of an echo against depth, 2αfd in decibels, with α ≈ 0.5 dB/(cm·MHz) for soft tissue.
  • A scanner with 60 dB of usable dynamic range can image until the round trip costs 60 dB: about 12 cm at 5 MHz, 30 cm at 2 MHz.
  • Higher frequency buys resolution: the readouts show the wavelength and the axial resolution for a two-cycle pulse.

Predict

Soft tissue has an acoustic impedance of 1.63 MRayl and air about 0.0004 MRayl. What fraction of ultrasound intensity reflects at a tissue–air boundary?

Name it

Impedance and echoes

A medium's acoustic impedance is Z=ρc; soft tissue is about 1060 kg/m³ × 1540 m/s = 1.63 MRayl, fat 1.38, bone about 7.8, air 0.0004. At a boundary hit squarely, the reflected fraction of intensity is

R=(Z2−Z1Z2+Z1)2

and 1−R goes on. It's the transmission-line reflection coefficient of C10, squared for power.

Ranging and timing

Scanners assume c = 1540 m/s everywhere, so an echo after time t comes from depth d=ct/2. The next pulse must wait for echoes from the deepest point, so the pulse repetition frequency is at most c/2dmax⁡, and an image of N lines takes N×2dmax⁡/c: the frame rate is c/(2dmax⁡N).

Attenuation

Soft tissue absorbs and scatters about 0.5 dB per cm per MHz, roughly proportional to frequency, so the round trip to depth d costs 2αfd dB. Time-gain compensation raises the receiver gain with time to even out the image.

Resolution

  • Axial (along the beam): half the spatial pulse length, nλ/2 for an n-cycle pulse, with λ = c/f.
  • Lateral: the beam width, about λ times the focal ratio (focal length over aperture) at the focus.

Both improve with frequency, which is limited by the depth needed: 2 to 5 MHz for abdomen and heart, 7 to 15 MHz for shallow structures.

Matching layers

PZT ceramic has an impedance of about 30 MRayl, so most energy would reflect at its face. A quarter-wavelength layer with Zm=ZPZTZtissue matches them, just as a quarter-wave transformer matches transmission lines.

Worked example

An echo returns after 130 µs. How deep was the reflector? How much is a 5 MHz echo from 10 cm attenuated, and what's the axial resolution of a two-cycle 5 MHz pulse?

Show the solution
  1. Depth. The echo travels there and back: d=1540×130×10−62m=10.0cm
  2. Attenuation. The round trip costs α=2×0.5×5×10=50dB
  3. Wavelength. λ=15405×106m=0.308mm
  4. Axial resolution. Δz=2×0.3082mm=0.31mm

Answer the reflector was 10.0 cm deep, the echo is attenuated by 50 dB, and the axial resolution is 0.31 mm.

Worked example

What frame rate does a 128-line image to 15 cm allow? And what impedance and thickness should a 5 MHz matching layer have, if sound travels at 3000 m/s in it?

Show the solution
  1. Time per line. t=2×0.151540s=195μs
  2. Frame rate. 128 lines take 25 ms, so the frame rate is 40 frames per second.
  3. Matching-layer impedance. Z=30×1.63=7.0MRayl
  4. Matching-layer thickness. A quarter wavelength in the layer: λ4=30005×106×4m=0.15mm

Answer the image allows 40 frames per second, and the matching layer should be 7.0 MRayl and 0.15 mm thick.

Worked example

An unfocused 5 MHz transducer is 10 mm across. How long is its near field, how fast does its beam spread beyond it, and what lateral resolution does focusing at f-number 5 give?

Show the solution
  1. Wavelength. In soft tissue: λ=15405×106m=0.308mm
  2. Near field. The beam stays about as wide as the transducer for N=D24λ=81mm
  3. Spread beyond it. The beam spreads at a half-angle of θ=sin−1⁡1.22λD=2.2°
  4. Focused lateral resolution. Focusing narrows the beam at the focus to about λ times the f-number: 0.308×5=1.5mm but only over a limited depth, which is why scanners focus electronically at several depths.

Answer the near field is 81 mm long, the beam spreads at 2.2° beyond it, and focusing at f-number 5 gives about 1.5 mm of lateral resolution.

Check

Ultrasound passes from a medium of 1.63 MRayl into one of 1.99 MRayl. What fraction of the intensity reflects?

Check

An echo arrives 65 µs after the pulse. How deep is the reflector?

Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.

Check

Tissue attenuates by 0.5 dB/(cm·MHz). How much is a 2 MHz echo from 15 cm deep attenuated?

Check

What's the axial resolution of a 2-cycle pulse at 2 MHz in soft tissue?

Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.

Check

An image of 64 lines reaches 15 cm deep. What's the highest possible frame rate, in frames per second?

Check

What impedance, in MRayl, should a quarter-wave layer have to match a 30 MRayl transducer to 1.5 MRayl tissue?

Check

An ultrasound transducer 6 mm across works at 5 MHz in soft tissue (1540 m/s).

(a)

How long is its near field?

Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.

(b)

What lateral resolution does focusing at f-number 4 give?

Check

Sound travels at 1540 m/s in tissue.

Step 1

How deep, in centimeters, is a reflector whose echo returns after t µs?

Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.

Explain why

Why must an ultrasound system trade imaging depth against resolution?

Practice

Keep practicing

10 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.