S15 Biomedical instrumentation · Practice
Ultrasound imaging
Problems dealt one at a time, with new numbers every time. The kinds you miss come back more often, and you can keep going as long as you like.
0 solved
0 in a row
0 of 10 kinds solid
Check
Ultrasound passes from a medium of 1.63 MRayl into one of 1.99 MRayl. What fraction of the intensity reflects?
Check
An echo arrives 65 µs after the pulse. How deep is the reflector?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
Check
Tissue attenuates by 0.5 dB/(cm·MHz). How much is a 2 MHz echo from 15 cm deep attenuated?
Check
What's the axial resolution of a 2-cycle pulse at 2 MHz in soft tissue?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
Check
An image of 64 lines reaches 15 cm deep. What's the highest possible frame rate, in frames per second?
Check
What impedance, in MRayl, should a quarter-wave layer have to match a 30 MRayl transducer to 1.5 MRayl tissue?
Check
An ultrasound transducer 6 mm across works at 5 MHz in soft tissue (1540 m/s).
(a)
How long is its near field?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
(b)
What lateral resolution does focusing at f-number 4 give?
Check
Intensity reflection: 0.0004 → 1 MRayl?
Check
Round-trip loss: α = 0.3, 1 MHz, 25 cm?
Check
Frame rate: 2 cm, 32 lines?
A kind of problem counts as solid once you've got it right on the first try twice in a row. Back to the lesson · Practice every topic in S15