S15 Biomedical instrumentation · Practice
Capacitive, inductive and piezoelectric sensors
Problems dealt one at a time, with new numbers every time. The kinds you miss come back more often, and you can keep going as long as you like.
Check
A capacitive sensor has 1 cm² plates 0.05 mm apart with relative permittivity 4. What's its capacitance?
Write units the way you normally would: 1.5 pF and 0.0015 nF are the same answer.
Check
A 20 pF parallel-plate sensor has a 50 µm gap. How much does its capacitance change when the gap shrinks by 2 µm?
Write units the way you normally would: 1.5 fF and 0.0015 pF are the same answer.
Check
A piezoelectric sensor with d = 150 pC/N and 10 nF of capacitance is suddenly loaded with 0.5 N. What voltage appears at first?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
Check
A piezoelectric sensor of 0.5 nF is read across 1 GΩ. What's its low-frequency corner?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
A piezoelectric sensor with d = 100 pC/N drives a charge amplifier with = 1000 pF. What's the size of the output for a 0.2 N force step?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
Check
A piezoelectric sensor jumps to 1 V under a steady load and is read across 2 GΩ with 2 nF of capacitance. What's left after 0.2 s?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
Check
A differential capacitor has 1 cm² plates, each 50 µm from a center plate in air, which moves 5 µm.
(a)
What is ?
Write units the way you normally would: 1.5 pF and 0.0015 nF are the same answer.
(b)
What is ?
Check
C: 0.1 cm², 0.01 mm, εr = 10?
Write units the way you normally would: 1.5 pF and 0.0015 nF are the same answer.
Check
Initial voltage: d = 2 pC/N, 0.1 N, 20 nF?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
Check
Corner: 0.01 GΩ, 0.1 nF?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
A kind of problem counts as solid once you've got it right on the first try twice in a row. Back to the lesson · Practice every topic in S15